22
i.
Run the program listed in step e. The value displayed by the controller should be the value noted in step h:
±
0.02Vdc (6033A).
±
0.092Vdc (6038A).
Load Effect (Load Regulation). Constant-voltage load effect is the change in dc output voltage (Eo) resulting from a
load-resistance change from open-circuit to full-load. Full-load is the resistance which draws the maximum rated output
current at voltage Eo. Proceed as follows:
a.
Connect the test equipment as shown in Figure 2-5. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b.
Turn the unit’s power on, and, using DISPLAY SETTINGS pushbutton switch, turn up current setting to full output.
c.
Turn up output voltage to:
7.0Vdc (6033A).
20.0Vdc (6038A).
as read on the digital voltmeter.
Figure 2-5. Basic Test Setup
d.
Reduce the resistance of the load to draw an output current of:
29Adc (6033A).
10 Adc (6038A).
Check that the unit’s CV LED remains lighted.
e.
Open-circuit the load.
f.
Record the output voltage at the digital voltmeter.
g.
Reconnect the load.
h.
When the reading settles, record the output voltage again. Check that the two recorded readings differ no more than:
±
0.0027Vdc (6033A).
±
0.005Vdc (6038A).
Source Effect (Line Regulation). Source effect is the change in dc output voltage resulting from a change in ac input
voltage from the minimum to the maximum value as specified in Input Power Requirements in the Specifications Table, in
the Operating Manual. Proceed as follows:
Summary of Contents for 6038A
Page 4: ......
Page 8: ......
Page 34: ...34 Figure 3 1 Troubleshooting Isolation ...
Page 35: ...35 Figure 3 1 Troubleshooting Isolation continued ...
Page 37: ...37 Figure 3 3 Clock and Primary SA Waveforms ...
Page 40: ...40 Figure 3 4 Readback and Secondary SA Waveforms ...
Page 56: ...56 Figure 3 7 Waveforms ...
Page 62: ...62 Figure 4 1 HP IB Block Diagram ...
Page 65: ...65 Figure 4 2 Front Panel Block Diagram ...
Page 90: ......
Page 93: ...93 Figure 6 1 Top View Top Covers Removed ...
Page 94: ...94 Figure 6 2 Main Board A1 Component Location ...
Page 95: ...95 Figure 6 3 Control Board A2 Component Location ...
Page 96: ...96 Figure 6 4 Front Panel Board A3 Component Location ...
Page 97: ...97 Figure 6 5 Power Mesh Board A4 Component Location ...
Page 98: ...98 Figure 6 6 HP IB Board A8 Component Location ...
Page 105: ...105 Figure 3 1 Troubleshooting Isolation option 001 ...
Page 106: ......