Softkeys and Displays
AUTO ATTEN
AUTO ATTEN
This softkey triggers a one-time autoranging of the downconverter’s input
attenuator.
The downconverter autoranges by first setting the attenuator to its maximum
attenuation (35 dB) then reducing the attenuation in 5 dB steps until the
mixer’s (or RF AM detector’s) input level is between 0 and 5 dB below the
limit shown on the
Mixer Max Level:
(or
AM DET MAX Lvl:
) display. The
level is measured by the downconverter’s internal voltmeter. If no signal is
measured when the attenuation reaches 0 dB, the attenuator is reset to 35 dB.
Path
Amplitudes more 1 of 2 AUTO ATTEN
or,
AM Noise AUTO ATTEN
HP-IB command
DC:AMPL:INP:ATT AUTO
Figure 10-31
HP 70427A Downconverter
Summary of Contents for 70427A