![background image](http://html.mh-extra.com/html/hp/70427a/70427a_user-manual_157664077.webp)
2-32
HP 70427A/HP 70428A User’s Guide
Getting Started
To Output a Signal with a Microwave Source
Step 2 - Set the output power
A frequency must be shown in the
Frequency:
display before an output
power entry will be accepted.
1. Press
the
OUTPUT POWER
softkey.
2. Press
the
POWER
softkey.
3. Key in the output power, dBm or mW, with the numeric keypad.
Power levels between 0 and +20 (100 mW), in 0.1 dB steps, can be
entered. But the minimum specified power is shown below.
•
2.4 to 6.6 GHz:+16
•
7.2 to 25.8 GHz: +10
Also, the actual resolution can be as much as 0.2 dB.
4. Press
the
dBm
or
mW
softkey.
Figure 2-34
Output Power Softkey Locations