*IDN?
returns the following response strings depending on model and options:
HEWLETT-PACKARD,E1413C
,
<
serial number
>
,
<
revision number
>
HEWLETT_PACKARD,E1313A_32CH
,
<
serial number
>
,
<
revision number
>
HEWLETT_PACKARD,E1313A_64CH
,
<
serial number
>
,
<
revision number
>
•
The C-SCPI type for this returned value is string.
Note
The revision will vary with the revision of the driver software installed in your
system. This is the only indication of which version of the driver is installed.
*LMC?
Learn Macros Query. Returns a quoted string name for each currently defined
macro. If more than one macro is defined, the strings are separated by commas (,).
If no macro is defined,
*LMC?
returns a null string.
*OPC
Operation Complete. Causes an instrument to set bit 0 (Operation Complete
Message) in the Standard Event Status Register when all pending operations have
been completed. By enabling this bit to be reflected in the Status Byte Register
(
*ESE 1
command), you can ensure synchronization between the instrument and an
external computer or between multiple instruments.
Note
Do not use
*OPC
to determine when the
CALibration:SETup or CALibration:TARE
commands have completed. Instead, use their query forms
CALibration:SETup?
or
CALibration:TARE?
.
*OPC?
Operation Complete Query. Causes an instrument to place a 1 into the
instrument’s output queue when all pending instrument operations are finished. By
requiring your computer to read this response before continuing program execution,
you can ensure synchronization between one or more instruments and the computer.
The C-SCPI type for this returned value is int16.
Common Command Reference
278 HP E1313/E1413 Command Reference