type
>
discrete
numeric
numeric
TC: B, E, J, K, N14, N28, R, S, or T
THER/FTH:2252 | 5000 | 10000
RTD/FRTD: 85 | 92
none
ohms
alpha
<
channel_list
>
numeric
Refer to the multiplexer user’s manual for
a list of channels available.
none
Comments
•
The HP E1344A, E1347A, E1353A, or E1476A multiplexer is required for
thermocouple measurements. These module’s have built-in thermocouple
compensation circuitry on the terminal module.
•
The channel list is of the form
(@ccnn)
,
(@ccnn,ccnn)
, or
(@ccnn:ccnn)
, where
cc
= card number and
nn
= channel number (105 is channel 05 of card number 1,
for example).
•
Four-wire temperature measurements use channel pairs. For example, on the HP
E1345A multiplexer, channels 0 and 8, 1 and 9, 2 and 10, etc. are paired. The lower
channel in each pair (0, 1, 2, ...7) is the sense channel. Use channel_list to specify the
"sense" channels.
•
You can measure RTD types 85 (alpha = 0.00385
Ω/Ω/°
C)
and 92 (alpha = 0.00392
Ω/Ω/°
C). The values 385, 0.00385, 392, 0.00392 are
also accepted for the type parameter.
•
Thermistor types are 2252, 5000, and 10000. Use thermistors that match the
Omega 440xx series temperature response curves.
•
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making Thermocouple Measurements
CONF:TEMP TC,J,(@100:107)
!Measure J-type thermocouples (scan the 8
channels once); trigger source is IMMediate
by default.
READ?
!Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement
!Enter readings into computer.
:VOLTage:AC
CONFigure:VOLTage:AC [<
range
>[,<
resolution
>]] [,<
channel_list
>]
selects the
AC-coupled RMS voltage function and allows you to specify the range and
resolution. If you specify a channel list, those multiplexer channels are scanned.
For a complete listing of range and resolution values available, see
Table 5-1 on page 120.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
0.0795 V | 0.63V | 5.09 V |
40.7 V | 300 V | AUTO | DEF |
MIN | MAX
volts
Chapter 5
Command Reference 130
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com