OUTPut
The
OUTPut
command subsystem enables you to route the multimeter’s voltmeter
complete signal to the VXIbus TTL trigger lines.
Subsystem Syntax
OUTPut
:TTLTrgn[:STATe] <mode>
:TTLTrgn[:STATe]?
:TTLTrg
n
[:STATe]
OUTPut:TTLTrg
n
[:STATe] <
mode
>
enables or disables routing of the voltmeter
complete signal to the specified VXIbus trigger line (TTLTrg0 through TTLTrg7)
on the backplane P2 connector.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
n
discrete
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7
none
<
mode
>
boolean
OFF | 0 | ON | 1
none
Comments
•
The
OUTPut
subsystem applies to the HP E1411B multimeter only.
•
You can substitute decimal values for the
OFF
("
0
") and
ON
("
1
") parameters.
•
The voltmeter complete signal is always routed to the E1411B multimeter’s front
panel "VM Complete" BNC connector. When enabled (
ON
), the
OUTPut
command also routes voltmeter complete to the specified trigger line on connector
P2. When disabled (
OFF
), voltmeter complete is routed only to the multimeter’s
front panel connector.
•
The multimeter generates the voltmeter complete signal after it has sampled the input
for each reading. The length of time this low-going TTL signal is true (low) depends
on the aperture time and on the autozero mode as shown below.
Aperture Time
Voltmeter Complete low
Autozero ON Autozero OFF
320 ms (50 Hz)
267 ms (60 Hz)
20 ms (50 Hz)
16.7 ms (60 Hz)
2.5 ms (400 Hz)
100
µ
s
10
µ
s
350 ms
370
µ
s
20.5 ms
17.2 ms
3.1 ms
520
µ
s
NA
350
µ
s
370
µ
s
370
µ
s
390
µ
s
430
µ
s
250
µ
s
70
µ
s
•
The VXIbus trigger lines are open-collector TTL lines that remain in a
non-asserted (high) state until the voltmeter complete signal is sent.
•
*RST Condition:
OUTP:TTLT
n
OFF
153 Command Reference
Chapter 5