SAMPle
The
SAMPle
command subsystem operates with the
TRIGger
command subsystem.
The
SAMPle
subsystem:
Designates the number of readings made for each trigger signal received
(
SAMPle:COUNt
).
Selects the pacing source for the sample period (
SAMPle:SOURce
).
Sets the sample period when the sample count is greater than one (
SAMPle:TIMer
).
Subsystem Syntax
SAMPle
:COUNt <number>
:COUNt? [MIN | MAX]
:SOURce <source>
:SOURce?
:TIMer <period>
:TIMer? [MIN | MAX]
:COUNt
SAMPle:COUNt <
number
>
designates the number of readings per trigger.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
1 - 16,777,215 | MIN | MAX
none
Comments
•
MINimum
sets 1 reading per trigger.
MAXimum
sets 16,777,215 readings per
trigger.
•
If
MAX
or 16,777,215 is specified for number, an “Out of memory” error is
generated to show that memory is exceeded. However, you can execute
READ?
to return the readings to the output buffer.
•
CONFigure
and
MEASure
set the sample count to 1.
•
*RST Condition:
SAMP:COUN 1
Example
Setting the Sample Count
CONF:VOLT:DC
!Function: DC voltage; stand-alone
multimeter.
TRIG:SOUR EXT
!Trigger source is external BNC on
multimeter front panel.
SAMP:COUN 10
!Specify 10 readings per trigger.