:DELay:AUTO
TRIGger:DELay:AUTO <
mode
>
enables or disables a trigger delay based on the
present function, range, and integration time. The trigger delay specifies the period
between the trigger signal and the start of the measurement.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
mode
>
boolean
OFF | 0 | ON | 1
none
Comments
•
You can substitute decimal values for the
OFF
("
0
") and
ON
("
1
") parameters.
•
When
TRIG:DEL:AUTO
is
ON
, the trigger delay is 0 seconds for DC voltage and
resistance measurements or 0.5 seconds for AC voltage measurements.
•
For the stand-alone multimeter, the trigger delay is inserted between the trigger
and the first measurement of a burst. For the scanning multimeter, the trigger
delay is inserted between the trigger and the first channel in each scan.
•
If a trigger delay is specified using the
TRIGger:DELay <
period
>
command,
TRIG:DEL:AUTO
is turned
OFF
.
•
You can set a delay between measurements in a burst using the
SAMPle:TIMer
command.
•
*RST Condition:
TRIG:DEL:AUTO ON
Example
Disabling Automatic Trigger Delay
TRIG:DEL:AUTO OFF
!Disable automatic trigger delay.
:DELay:AUTO?
TRIGger:DELay:AUTO?
returns a number to show whether the automatic trigger
delay mode is on or off: "
1
" =
ON
, "
0
" =
OFF
. The number is sent to the output
buffer.
Example
Querying the Trigger Delay Mode
TRIG:DEL:AUTO OFF
!Disable automatic trigger delay.
TRIG:DEL:AUTO?
!Query multimeter to return trigger delay
mode.
enter statement
!Enter value into computer.
Chapter 5