Four-Wire Resistance
Input Characteristics
Measurement Characteristics vs. Range
Range
Source
Current
Maximum
Open
Circuit
Voltage
Maximum
Allowable
Current
Source
Lead
Resistance
Maximum
Allowable
Offset Volts
For Offset
Compensated
Ohms
Default
Measurement
Settling
Time
Default
Range/
Function
Change
Settling
Time
256
Ω
488
µ
A
11.5 V
20 k
Ω
12 mV
0
0
2 k
Ω
488
µ
A
11.5 V
15 k
Ω
100 mV
0
0
16 k
Ω
61
µ
A
11.5 V
100 k
Ω
100 mV
0
0
131 k
Ω
61
µ
A
11.5 V
30 k
Ω
0.8
V
0 2
ms
1 M
Ω
7.6
µ
A
11.5 V
200 k
Ω
0.8
V
0 11
ms
Conditions: Settling times may need to be increased (programmably) if load capacitance is greater than 200 pF.
Resolution vs. Aperture (ohms)
Range
267/320 ms
16.7/20 ms
2.5 ms
100
µ
s
10
µ
s
256
Ω
.06 m
Ω
0.25 m
Ω
1.0 m
Ω
8.0 m
Ω
15 m
Ω
2 k
Ω
.5 m
Ω
2.0 m
Ω
8.0 m
Ω
60 m
Ω
125 m
Ω
16 k
Ω
4.0 m
Ω
15 m
Ω
60 m
Ω
500 m
Ω
1.0
Ω
131 k
Ω
30.0 m
Ω
125 m
Ω
500 m
Ω
4.0
Ω
8.0
Ω
1 M
Ω
.25 m
Ω
1.0
Ω
4.0
Ω
30
Ω
64
Ω
Accuracy Conditions for table below: Autozero on, one hour warmup. Within 24 hours
and
±
1
°
C of calibration temperature (module calibrated at 18
°
C to 28
°
C).
24-Hour Accuracy vs. Aperture
±
(% of r ohms)
Range
267/320 ms
16.7/20 ms
2.5 ms
100
µ
s
10
µ
s
256
Ω
.015% + 10 m
Ω
.015% + 10 m
Ω
.015% + 10 m
Ω
.05% + 50 m
Ω
.05% + 50 m
Ω
2 k
Ω
.015% + 20 m
Ω
.015% + 20 m
Ω
.015% + 20 m
Ω
.05% + 150 m
Ω
,
.05% + 200 m
Ω
16 k
Ω
.015% + 200 m
Ω
.015% + 200 m
Ω
.015% + 200 m
Ω
.05% + 1
Ω
.05% + 2
Ω
131 k
Ω
.015% + 1
Ω
.015% + 1
Ω
.015% + 1
Ω
.05% + 8
Ω
.05% + 16
Ω
1 M
Ω
.015% + 10
Ω
.015% + 10
Ω
.015% + 10
Ω
.05% + 60
Ω
.05% + 120
Ω
192 HP E1326B/E1411B Multimeter Specifications
Appendix A