If you have an HP 75000 Series B mainframe with a keyboard, errors
can be monitored by selecting "Monitor" from the multimeter menu.
If errors occur when the program executes, the "err" annunciator will
appear. Entering
SYST:ERR?
repeatedly from the keyboard reads
all of the messages in the error queue.
•
An overload condition (for example, reading = +9.037)
sets the Device Dependent Error bit in the Standard Event Status
Register. In this program, this condition interrupts the computer
which then calls the subprogram. However, an overload does not
generate an error message so 0 "No Error" is displayed.
•
The B-size VXIbus Mainframe User’s Manual contains detailed
information on the Status and Standard Event Status Registers.