Using MEASure and CONFigure Commands
Each time the multimeter makes a measurement, it does so from a
configuration based on several parameters. The easiest way to set these
parameters is with the
MEASure
and
CONFigure
commands:
MEASure:
measurement
?
[range|AUTO|DEF|MIN|MAX[,resolution|DEF|MIN|MAX]][,(@channel_list)]
CONFigure:
measurement
[range|AUTO|DEF|MIN|MAX[,resolution|DEF|MIN|MAX]][,(@channel_list)]
MEASure:TEMPerature? transducer,type[,(@channel_list)]
CONFigure:TEMPerature transducer,type[,(@channel_list)]
Executing these high-level commands is equivalent to setting up the
multimeter with the commands shown in Table 4-1 on page 77. Note that
specifying a channel list identifies a scanning multimeter. No channel list
identifies a stand-alone multimeter.
Note
If a channel list is the only parameter specified in the
MEASure
or
CONFigure
command, it must be separated from the command header by a
space, rather than a comma (e.g.
MEAS:VOLT:DC? (@100)
).
76 Understanding the HP E1326B/E1411B Multimeter