Table 4-1. Configurations Using MEASure and CONFigure
Parameter
Command
Setting
Function
As specified.
Range
VOLTage:RANGe
RESistance:RANGe
As specified or autorange.
Resolution
VOLTage:RESolution
RESistance:RESolution
As specified or a function of range and aperture or
integration time.
Aperture Time
VOLTage:APERture
RESistance:APERture
16.7 ms (60 Hz), 20 ms (50 Hz), or based on the
specified resolution.
Integration Time
VOLTage:NPLC
RESistance:NPLC
1 power line cycle (PLC) or based on the specified
resolution.
Autozero
CALibration:ZERO:AUTO
ON
; autozero is performed after every measurement.
Offset Compensation
RESistance:OCOMpensated
OFF
; resistance measurements only.
Trigger Source
TRIGger:SOURce
IMM
; trigger signal is always true. Measurement is taken
when multimeter goes to Trigger State.
Number of Triggers
or Number of Scans
TRIGger:COUNt
1
; number of triggers issued or number of scans through
channel list before multimeter returns to Idle State.
Trigger Delay
TRIGger:DELay
AUTO
- delays are 0 seconds for DC voltage and
resistance, 0.5 seconds for AC voltage.
Readings per Trigger
SAMPle:COUNt
1
; number of measurements taken when trigger is
received.
Sample Period
SAMPle:SOURce
IMM
; period between measurements or the period
between FET multiplexer scans.
Chapter 4
Understanding the HP E1326B/E1411B Multimeter 77