General Characteristics
Settling Time:
300
µ
S (+ full scale to – full scale step, single channel,
to accuracy listed above)
Isolation (isolated channels):
42 Vdc/42 V peak (channel-to-chassis or channel-to-channel)
Synchronization:
Software commands, External TTL trigger, or VXIbus TTL Trigger lines
Single channel or multiple channel
VXI Specification:
Device Type: A16 or A24, D16, slave only, register-based
Power Requirements
Current Required (Amps)
Configuration
+5 V
DC
+5 V
Dynamic
+12 V
DC
+12 V
Dynamic
+24 V
DC
+24 V
Dynamic
–24 V
DC
–24 V
Dynamic
8 Channel
Non-Isolated
0.70
0.01
0.04
0.01
0.44
0.01
0.44
0.01
8 Channel
Isolated
0.75
0.01
0.04
0.01
0.88
0.01
0.00
0.00
16 Channel
Non-Isolated
1.25
0.01
0.04
0.01
0.88
0.01
0.88
0.01
16 Channel
Isolated
1.40
0.01
0.04
0.01
1.60
0.01
0.00
0.00
Cooling Requirements
For a 10
°
C rise
Air Flow
∆
P
Avg Watts/slot
8 Channel configurations
2.0 liters/second
0.10 mm H
2
O
25.4
16 Channel configurations
3.9 liters/second
0.18 mm H
2
O
49.0
96 HP E1418A Specifications
Appendix A
Summary of Contents for E1418A
Page 4: ...Notes 4 HP E1418A 8 16 Channel D A Converter Service Manual Contents ...
Page 8: ...Notes 8 HP E1418A D A Converter Module Service Manual ...
Page 58: ...Notes 58 Verification Tests Chapter 4 ...
Page 66: ...Figure 5 5 Internal Resistor Adjustment Connections 66 Adjustments Chapter 5 ...
Page 81: ...Figure 6 1 8 Channel Disassembly Chapter 6 Service 81 ...
Page 82: ...Figure 6 2 16 Channel Disassembly 82 Service Chapter 6 ...
Page 84: ...Notes 84 Service Chapter 6 ...
Page 90: ...Figure 7 1 HP E1418A Replaceable Parts 90 Replaceable Parts Chapter 7 ...
Page 91: ...Figure 7 2 Common Terminal Module Replaceable Parts Chapter 7 Replaceable Parts 91 ...