APPLy Output
The
APPLy
n subsystem commands provide the easiest method to output
voltages or currents.
/* Set channel 1 for 5.0 Volt output */
APPLy1:VOLTage 5.000
/* Channel is set to voltage output and 5.000 V applied */
/* Set channel 8 for 10 mA output */
APPLy8:CURRent 0.0100
/* Channel is set to current output and 10 mA applied */
The
APPLy
n subsystem outputs the voltage or current specified on the
channel specified immediately. The output relay for that channel is closed.
Other channel outputs and output states are not disturbed.
An
APPLy
n subsystem command performs several actions with a single
command. Each
APPLy
n command affects a single channel and:
–
Sets the channel output mode (voltage or current).
–
Sets the output value (volts or amps).
–
Enables the outputs (closes the output relay).
The output channel remains enabled, and outputs the last programmed
voltage or current value, unless you take explicit action to turn off the
output. For example, if you send two
APPLy
n commands to the same
channel, the first command sets an output value and closes the output relay.
The second command changes only the output value (since the relay is
already closed).
/* Set channel 1 for 5.0 Volt output */
APPLy1:VOLTage 5.000
/* Channel 1 is set to voltage output and 5.000 V applied */
/* Now set channel 1 for 6.0 Volt output */
APPLy1:VOLTage 6.000
/* Channel 1 is set to voltage output and 6.000 V applied */
The commands above will set a 5.000 Volt output on Channel 1, and then
set a 6.000 Volt output on channel 1. The output relay is not affected by the
second command.
Chapter 2
Programming Examples 45
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...