Reset and Registers
Following power-on or a
*RST
command, the registers are put into the
following states:
–
The Manufacturer ID and Device ID Registers remain unaffected.
–
The Channel Mode Register (base + 1A
16
) is set to voltage output
for all channels (FFFF
16
) or is set to the state set by the V/I jumper
(described on page 31) if the P/J jumper is in the jumper selectable
position.
–
The Channel Program Jumper Register (base + 16
16
) is set to program
selectable for all channels (FFFF
16
) or to the position of the P/J jumper
–
The Channel Relay Control Register (base + 1C
16
) is set to open all
channel relays (FFFF
16
).
–
The Isolation Status Register (base + 14
16
) reflects the isolated/
non-isolated configuration of each channel.
–
The Calibration Control Register (base +08
16
) is set to calibrate
voltage with the cal mux disabled (2000
16
).
–
Channel Trigger Register (base + 18
16
) is set to 0000
16
if the
module is a 16-channel configuration or FF00
16
if an 8-channel
configuration.
Note
The upper 8-bits of all channel related registers are set to 1’s (FF
16
) for
8-channel configurations.
121 HP E1418A Register-Based Programming
Appendix B
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...