SOURce
n
:CURRent[:LEVel]:TRIGgered[:AMPLitude]
SOURce
n
:CURRent[:LEVel]:TRIGgered[:AMPLitude] <
value
| MIN | MAX | DEF>
outputs the current set in value at channel n when the trigger conditions are satisfied and
the output relay is closed. Use this command to synchronize multiple channel outputs.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n
keyword
1 to 8 or 1 to 16
1
<
value
>
numeric (float 32) or
discrete
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
0
Comments
•
In the keyword,
SOURce
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
The
<
value
>
parameter is a floating point 32-bit number in the range of -0.02 A
to +0.02 A.
MIN
sets -0.02 Amps.
MAX
sets +0.02 Amps.
DEF
sets 0 Amps.
•
This command sets channel n to output the current specified following a valid
trigger condition. Following this command, the trigger system must be enabled
using an
INITiate
command. See the
TRIGger
subsystem discussion on page 54
for additional information.
•
If the module is in a wait-for-trigger state (
INIT
has been received) when this
command is received, an error is generated.
•
The output state of the channel must be enabled and the trigger event satisfied for
the current to appear at the output terminals.
•
If the channel is in the voltage mode when this command is received, an error is
generated.
•
Related Commands:
SOURce
n
:CURRent[:LEVel]:TRIGgered[:AMPLitude]?
,
SOURce
n
:FUNCtion:MODE CURRent
,
OUTPut
n
[:STATe]
SOURce
Chapter 3
HP E1418A SCPI Command Reference 86
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...