32 Chapter 4 General procedures
ENWW
Calibrating the electronic scale
Calibrate the electronic scale once a day.
1
Connect the scale to a power source, and press the
On
button. Allow a one-minute
warm up period.
Note
Do
not
put the adjusting weight on the scale.
Adjusting weight and scale
2
Press and hold the
Cal/Mode
button until the exact size (
500.0
) of the adjusting
weight appears flashing on the LCD display.
Note
It may take up to two seconds for the scale to show results.
3
Press the
Off
button. The message
CAL 0
appears on the display.
4
Place the adjusting weight in the center of the weighing pan. After a short while, the
message
CAL F
appears on the display for a short time.
The exact weight of the adjusting weight (that is,
500.0
) should appear on the display
shortly afterwards.
If an adjusting error occurs or the wrong adjusting weight is used, the message
CAL E
appears on the display. In this case, repeat this calibration procedure.
5
Put the 500 g adjusting weight in a safe place.
Adjusting
weight
Weighing
pan
LCD
display
Cal/Mode
button
On button
Off button