B 215 / B 225 Evo
Operator’s manual
117
UK
2. The machine will display the following
message:
• CAL on the left-hand display;
• GO on the right-hand display.
3. Perform a first spin.
4. At the end of the spin, fix the calibration
weight supplied with the machine on the
shaft assembly bell as indicated in figure
23.
5. Perform a second spin.
6. At the end of the spin, change the position
of the calibration weight on the swinging
unit bell as indicated in figure 24.
7. Perform a third spin. This last calibration
phase includes the execution of three
consecutive spins in automatic mode.
At the end of the spin(the message "End CAL" will appear on the display), if the cali-
bration has been successful, an acoustic consent signal goes off at the end of the spin
otherwise the message E 2 is temporarily displayed.
NOTES
- Once the procedure is finished, remove the calibration weight.
- Pressing the key
, it is possible to interrupt the programme at any time.
- THIS CALIBRATION IS VALID FOR ANY TYPE OF WHEEL!
17. EXTERNAL MECHANICAL SENSOR
CALIBRATION (IF INSTALLED)
This programme is used to calibrate the width sensor potentiometer. It must be performed
when the machine requests it by displaying the E4 message, or when the measured and
the actual rim widths differ.
Proceed as follows to recall this programme:
• press the
key
• use the
keys to search for GAU CAL
• press the
key to perform calibration, proceeding as follows:
1. The machine will display the following message:
• CAL on the left display
24