background image

 

 

HT61 - HT62 

 

 

EN - 4

2. GENERAL DESCRIPTION 

The instrument carries out the following measurements: 

 

 DC 

Voltage 

 AC 

TRMS 

voltage 

  DC/AC voltage with low impedance (LoZ) 

 DC 

Current 

 AC 

TRMS 

Current 

  Resistance and Continuity test 

 Diode 

test 

 Capacitance 

(HT62) 

  Current and voltage frequency 

 Duty 

Cycle 

  Temperature with K probe (HT62) 

 
Each of these functions can be selected by means of the appropriate switch. The 
instrument is also equipped with function keys (see § 4.2), an analogue bargraph and 
backlight. The instrument is also equipped with an Auto Power OFF function (which can be 
disabled), which automatically switches off the instrument 15 minutes after the last time a 
function key was pressed or the rotary switch was turned. To switch on the instrument 
again, turn the rotary switch. 
 

2.1. MEASURING 

AVERAGE 

VALUES ANDTRMS VALUES 

Measuring instruments of alternating quantities are divided into two big families: 

  AVERAGE-VALUE meters: instruments measuring the value of the sole wave at 

fundamental frequency (50 or 60 Hz). 

  TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMS 

value of the quantity being tested. 

With a perfectly sinusoidal wave, the two families of instruments provide identical results. 
With distorted waves, instead, the readings shall differ. Average-value meters provide the 
RMS value of the sole fundamental wave; TRMS meters, instead, provide the RMS value 
of the whole wave, including harmonics (within the instruments bandwidth). Therefore, by 
measuring the same quantity with instruments from both families, the values obtained are 
identical only if the wave is perfectly sinusoidal. In case it is distorted, TRMS meters shall 
provide higher values than the values read by average-value meters. 
 

2.2.  DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR 

The root mean square value of current is defined as follows: 

In a time equal to a period, 

an alternating current with a root mean square value of 1A intensity, circulating on a 
resistor, dissipates the same energy that, during the same time, would be dissipated by a 
direct current with an intensity of 1A

". This definition results in the numeric expression: 

G=

T

t

t

dt

t

g

T

0

0

)

(

1

2

 The root mean square value is indicated with the acronym RMS

.

 

The Crest Factor is defined as the relationship between the Peak Value of a signal and its 

RMS value: CF (G)=

RMS

p

G

G

 This value changes with the signal waveform, for a purely 

sinusoidal wave it is 

2

=1.41. In case of distortion, the Crest Factor takes higher values 

as wave distortion increases. 

Summary of Contents for HT61

Page 1: ...Copyright HT ITALIA 2017 Version EN 2 00 17 07 2017 ENGLISH User manual...

Page 2: ...unction 8 5 OPERATING INSTRUCTIONS 9 5 1 DC Voltage measurement 9 5 2 AC Voltage measurement 10 5 3 AC DC Voltage measurement with low impedance LoZ 11 5 4 Frequency and Duty Cycle measurement 12 5 5...

Page 3: ...ment the following symbols are used Warning observe the instructions given in this manual improper use could damage the instrument or its components Double insulated meter AC voltage DC voltage or cur...

Page 4: ...alled overvoltage category is 6 7 4 Measured circuits reads OMISSIS Circuits are divided into the following measurement categories Measurement Category IV is for measurements performed at the source o...

Page 5: ...families of instruments provide identical results With distorted waves instead the readings shall differ Average value meters provide the RMS value of the sole fundamental wave TRMS meters instead pr...

Page 6: ...nt We also recommend checking that the packaging contains all components indicated in 6 3 1 In case of discrepancy please contact the Dealer In case the instrument should be returned please follow the...

Page 7: ...MENT CAPTION 1 LCD display 2 RANGE key 3 MAXMIN key 4 Hz key 5 REL key 6 MODE key 7 HOLD key 8 Rotary selector switch 9 Input terminal 10A 10 Input terminal VHz HT61 or Hz V HT62 11 Input terminal mA...

Page 8: ...e the same key is pressed again The display shows the symbol associated with the selected function MAX for maximum value and MIN for minimum value The MAX MIN key is not active when the HOLD function...

Page 9: ...arily connect the test leads between phase and neutral cables at least for 5s then perform the phase PE measurement to avoid unexpected trips out 4 2 8 Disabling the Auto Power Off function The instru...

Page 10: ...il the symbol DC is displayed 3 Insert the red cable into input terminal VHz HT61 or Hz V HT62 and the black cable into input terminal COM 4 Position the red lead and the black lead respectively in th...

Page 11: ...mbol AC is displayed 3 Insert the red cable into input terminal VHz HT61 or Hz V HT62 and the black cable into input terminal COM 4 Position the red lead and the black lead respectively in the spots o...

Page 12: ...ed cable into input terminal VHz HT61 or Hz V HT62 and the black cable into input terminal COM 4 Position the red lead and the black lead respectively in the desired spots of the circuit to be measure...

Page 13: ...ion Hz 2 Press the Hz key to select measurements Hz or in order to display the values of frequency and duty cycle of input voltage 3 Insert the red cable into input terminal VHz HT61 or Hz V HT62 and...

Page 14: ...or Hz V HT62 and the black cable into input terminal COM 3 Position the test leads in the desired spots of the circuit to be measured see Fig 6 The display shows the value of resistance 4 If the disp...

Page 15: ...red cable into input terminal VHz HT61 or Hz V HT62 and the black cable into input terminal COM 4 Position the leads at the ends of the diode to be tested see Fig 7 respecting the indicated polarity T...

Page 16: ...acitance measurement 1 Select position 2 Press the MODE key until the symbol nF is displayed 3 Insert the red cable into input terminal Hz V and the black cable into input terminal COM 4 Press the REL...

Page 17: ...C F 2 Press the MODE key until the symbol C or F is displayed 3 Insert the provided adapter into input terminals Hz V polarity and COM polarity see Fig 9 4 Connect the provided K type wire probe or t...

Page 18: ...sert the red cable into input terminal 10A or into input terminal mA A and the black cable into input terminal COM 4 Connect the red lead and the black lead in series to the circuit whose current you...

Page 19: ...ent 4 Insert the red cable into input terminal 10A or into input terminal mA A and the black cable into input terminal COM 5 Connect the red lead and the black lead in series to the circuit whose curr...

Page 20: ...sition to position and remove it 3 Remove the battery and insert a new battery of the same type see respecting the indicated polarity 4 Restore the battery compartment cover into place and turn the fa...

Page 21: ...up to 1000V DC AC TRMS voltage with low impedance LoZ Range Resolution Accuracy 50 400Hz Input impedance Protection against overcharge 600 0mV 0 1mV 3 0 rdg 40dgt approx 3k 600VDC ACrms 6 000V 0 001V...

Page 22: ...g 8dgt 1000VDC ACrms 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 0 001kHz 99 99kHz 0 01kHz 999 9kHz 0 1kHz 9 999MHz 0 001MHz 40 00MHz 0 01MHz Sensitivity 0 8Vrms 20 80 duty cycle and f 100kHz 5Vrms 20 80 du...

Page 23: ...nt backlight and bargraph Sampling frequency 2 times s 7 2 ENVIRONMENT 7 2 1 Environmental conditions for use Reference temperature 18 C 28 C 64 F 82 F Operating temperature 5 C 40 C 41 F 104 F Allowa...

Page 24: ...sequence of improper packaging Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel Modifications to the instrument performed without the manufactur...

Reviews: