XL421 - XL422 

 

 

6 MEASURING 

PROCEDURE 

6.1  USING THE INSTRUMENT IN A SINGLE PHASE PLANT 

 

 

WARNING 

 

 

 

The maximum overvoltage is CAT III 1000V

 to ground or CAT IV 600V

 to 

ground. Do not attempt to take any voltage measurements exceeding the 
limits indicated in this manual. Exceeding the limits could cause electrical 
shock or damage to the instrument. 

 

P

N

I1

P

N

I1

 

Fig. 2: Instrument’s connection in Single Phase plant 

 

WARNING 

 

 

If possible de-energize the plant under test before connecting the instrument. 
Take all the safety precautions before taking actions on the plant. 

 
1.  De-energize the plant under test before connecting the meter. 
2.  Open the flexible head by pressing on expected points (see Fig. 1). 
3.  Clamp the cable under measurement with the head and close it. 
4.  To increase the reading accuracy the clamped cable must be in the middle of the head. 
5.  Re-energize the plant under test. 
6. 

How to start a Recording:

 

  If no recording was performed or if the recording data was already downloaded to a 

PC, pressing and keeping the 

START/STOP

 key for 3 seconds. The STATUS LED 

will be lighting for the same time, than will be stable lighting for 1 second and the 
meter will start the recording. 

  If the recording data was not already downloaded to a PC 

but the operator wants 

to overwrite these data

, pressing and keeping the 

START/STOP

 key for 6 

seconds. The STATUS LED will be quickly blinking during the key pressing for the 
same time, than the LED will be stable lighting for 1 second and the instrument will 
start the recording. 

The old recording will be deleted and overwritten by the new 

one.

 The meter will wait for the next minute beginning (second=00) and the STATUS 

LED is double blinking. After that the recording starts and the STATUS LED blinks 
every 3 seconds. The instrument is storing the data. 

 

EN - 7

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for XL421

Page 1: ...ENGLISH User s manual Copyright HT ITALIA 2007 Release EN 1 02 13 06 2007 www com information itm com 1 800 561 8187...

Page 2: ...6 6 MEASURING PROCEDURE 7 6 1 Using the instrument in a Single Phase plant 7 6 2 Using the instrument in a Three Phase plant 9 7 LED MESSAGES DESCRIPTION 11 8 CONNECTION OF INSTRUMENT TO PC 11 9 MAIN...

Page 3: ...s Danger high voltage risk of electrical shocks De energize the plant under test before connecting the instrument Double insulated meter AC voltage and current 1 1 PRELIMINARY INSTRUCTIONS This instru...

Page 4: ...the source of low voltage installations Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units Measurement category III is for measurement...

Page 5: ...ingle phase or three phase plant Where not expressly indicated the characteristics are common for both models The meter can monitoring a plant for a long time thanks to a sophisticated memory managing...

Page 6: ...ph 10 1 2 For battery fitting operation procedure please see paragraph 9 2 WARNING The meter after fitted batteries is always continuously ON also if STATUS and ALARM LED are OFF Pressing of START STO...

Page 7: ...rnal clock Set instrument date and time Check battery status How to operate 1 Connect the instrument to PC COM using included serial cable 2 Run DATALINK software 3 Choose the correct COM port with th...

Page 8: ...n the middle of the head 5 Re energize the plant under test 6 How to start a Recording If no recording was performed or if the recording data was already downloaded to a PC pressing and keeping the ST...

Page 9: ...ion is possible During Recording operations RS 232 serial communications to PC is not possible 8 Stop a Recording To stop a Recording pressing START STOP key and keep it pressed for 3 seconds The STAT...

Page 10: ...he middle of the head 5 Re energize the plant under test 6 How to start a Recording If no recording was performed or if the recording data was already downloaded to a PC pressing and keeping the START...

Page 11: ...ion is possible During Recording operations RS 232 serial communications to PC is not possible 8 Stop a Recording To stop a Recording pressing START STOP key and keep it pressed for 3 seconds The STAT...

Page 12: ...NNECTION OF INSTRUMENT TO PC Connection of the instrument to a PC is made with a serial cable supplied with the instrument To download data to the PC after installation of DATALINK software keep to th...

Page 13: ...ING Only experts and trained technicians should perform this operation Remove the flexible head from the conductor under test before replacing the batteries 1 Stop the recording if necessary remove th...

Page 14: ...ht 2000m Overvoltage category CAT III 1000V to ground CAT IV 600V to ground 10 1 2 General data Mechanical characteristics Size 120 L x 80 W x43 D mm Max diameter cable clamped 174mm Head length 600mm...

Page 15: ...Damages caused by incorrect shipping procedures Damages caused by repair service trials of unauthorized people Modifications to the equipment without explicit authorization of our Technical Dept Adapt...

Reviews: