The electrode converter (options) is designed for the use of other electrodes, usually

moistureproof and dustproof.

When measuring, remain the wire of the electrode still, otherwise it will cause unstable

measurement.

The instrument used by power supply should be good grounding.

The instrument adopt MOS integrated circuit, so ensure the soldering iron have good

grounding during maintenance.

When using a buffer solution calibration instruments, to ensure the reliability of the

buffer solution, can't match wrong buffer solution, otherwise will lead to produce error
measurement results.

4.2 Usage and Maintenance of Electrode

The electrode must be calibrated with known standard buffer solution before operation,

and the pH of test solution should be closer to the pH buffer solution.

Remove the electrode cover, avoid the contact of electrode sensitive glass bulb with

hard objects, because any broken or abraded will make electrode invalid.

After measurement, put on protective cover timely. There should be a few reference

filling liquid in protective cover to keep the electrode bubble wet, must avoid immerse
in DI water.

The outer reference filling liquid of combination electrode is KCl solution for 3 mol/L.

Add fluid from the upper hole. When not in use, pull on the rubber case to prevent
liquid dry up.

Keep the electrode terminal clean and dry, absolutely prevent short circuit on both ends

of the output, otherwise will lead to misalignment or failure measurement.

To maintain good performance, match the electrode with pH meter of high input

impedance (

1012 Ω) .

Avoid long-term immersion in DI water, protein solution, acid and fluoride solution.

Avoid contact with the organic silicone oil low.

After long-term usage, if found that the slope is slightly lower: immerse the bottom of

electrode in 4% HF (HF) in (3 ~ 5) s, rinse with DI water, and then immerse in
0.1mol/L hydrochloric acid solution soak to renew electrode.

If there are substances that can pollute sensitive bulb or block liquid junction in test

solution, the electrode will be passivation and the slope is reduced, the display
reading no phenomenon. Users should use appropriate cleaning solution to renew
electrode according to the nature of pollutants,

Summary of Contents for PHS-2F

Page 1: ...PHS 2F Benchtop pH Meter Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...perate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrument I...

Page 3: ...Contents I Installation 1 II Specifications and Performance 2 III Operations 4 IV Maintenance and Service 8 V Packing List 11...

Page 4: ...PHS pH Combination Electrode 1pc 4 Electrode holder 1pc 1 2 Installation of the meter Front view of PHS 2F Rear panel of the meter 1 Meter 2 Keypad 3 Screen 4 Electrode Holder 5 PH pH Combination Ele...

Page 5: ...e BNC short circuit 11 plug on electrode socket 6 c Insert the plug of pH Combination Electrode 12 in pH electrode socket 6 II Specifications and Performance 2 1 Introduction Note following problems w...

Page 6: ...r more pH buffer solution 2 1 2 Features Adopt new design appearance big LCD segmented display It is mainly used to measure the pH value and potential value mV of water solution in the field of resear...

Page 7: ...mV measuring state it is cancel key when set temperature E0 and slope press this key to exit function module and back to measuring state E0 Press this key to adjust E0 value to increase to decrease S...

Page 8: ...ure of tested solution Then press Enter to confirm current temperature or press pH mV to back to measuring state The preparation of pH electrode 1 Remove the protective cover 13 of pH electrode and pu...

Page 9: ...n such as buffer solution pH 7 00 2 Measure the temperature of the solution with thermometer and set temperature value as previous method 3 After the reading becomes stable press E0 or E0 to adjust th...

Page 10: ...he instrument has been calibrated it can be used to measure test solution Whether the temperature of test solution and standard buffer solution are the same can cause different measurement procedure W...

Page 11: ...reference electrode into reference electrode socket 7 3 6 Turn off After operation press On Off to turn off the meter Immerse electrode into DI water when finish measurement If for long term not in u...

Page 12: ...e a few reference filling liquid in protective cover to keep the electrode bubble wet must avoid immerse in DI water The outer reference filling liquid of combination electrode is KCl solution for 3 m...

Page 13: ...ion in case of corrosive gas 4 If the meter still does not work well when the previous situations are eliminated contact the relevant departments 1 Do not choose cleaning solutions like carbon tetrach...

Page 14: ...11 V Packing List No Item Qty 1 pH combination electrode 1 2 Electrode Holder 1 3 Power Liner 1 4 pH Standard Buffer powder Each 5 5 Fuse 2 6 Protective Caver 1...

Reviews: