Pub. 42004-565B 

TempSense Model TS100 

Page 

2

 of 6

 

P:\Standard IOMs - Current Release\42004 Instr. Manuals\42004-565B.docx 

03/21 

Installation

 

N

OTE

:

  The accuracy of the TempSense TS100 requires individual subjects pass within two to six feet of 

the thermal imaging camera lens.  Results with different camera positions may vary.  Thermal 
imaging camera accuracy improves as the subject moves closer to the camera. 

Camera Location

 

Locate the TempSense TS100 so that people in a queue individually pass through the two cameras’ shared 
FOV (field of view) while facing the cameras.  Each subject scanned should come within two to six feet 
of the cameras.  The ideal height for camera installation is 5 feet 5 inches above the floor.  Installation at a 
higher location requires aiming the TS100 downward so that each subject’s forehead passes through the 
center of the FOV (see Figure 2 and Figure 3)

 

Figure 2.  SAMPLE FIGURE Camera Positioning (Side View) 

 

Figure 3.  SAMPLE FIGURE Camera Positioning (Top View) 

Summary of Contents for GAI-TRONICS TempSense TS100

Page 1: ...96 5954 VISIT WWW GAI TRONICS COM FOR PRODUCT LITERATURE AND MANUALS Confidentiality Notice 1 General Information 1 Installation 2 Camera Location 2 Wall Mount 3 Pole Mount 3 VESA Mount 3 Stand Mount...

Page 2: ...onics product or system This manual may not be disclosed in any form in whole or in part directly or indirectly to any third party General Information The TempSense Model TS100 measures and captures E...

Page 3: ...o the camera Camera Location Locate the TempSense TS100 so that people in a queue individually pass through the two cameras shared FOV field of view while facing the cameras Each subject scanned shoul...

Page 4: ...e the TS100 mounting plate see Figure 4 to a pole using customer supplied worm drive clamps 2 Secure the TS100 to the mounting plate using the four included 6 32 screws VESAMount Attach a VESA video e...

Page 5: ...100 s IR sensor has a white LED that blinks slowly for 40 seconds then changes to a fast blink NOTE If the IR sensor LED does not flash press the sensor into its socket using a soft cloth then reboot...

Page 6: ...nce from where the TS100 will scan subjects The calibration object must be the hottest object in the FOV 2 While watching the temperature of the object in the IR video feed left side press the setup b...

Page 7: ...splays a still image of the IR and standard camera s picture OVER TEMPERATURE flashes on screen notifying the operator of the overtemperature subject The TS100 displays the measured temperature of the...

Reviews: