Commissioning
OPERATION MANUAL
Chapter 2
Unichiller® eo Desktop
V2.2.0en/09.08.21//17.12
24
2.1.3
Positioning the temperature control unit
2.1.3.1
Temperature control unit with casters
▪
Do
not
use the casters for the transportation to the place of installation.
→
ing and transporting the temperature control unit«
▪
Use the casters only for positioning at the place of installation.
▪
Only ever move the temperature control unit on the casters if the surface is level, without gradi-
ent, non-slip and stable.
▪
Do not move the temperature control unit alone.
▪
At least 2 persons
are required to move the temperature control unit on casters.
At least 5 per-
sons
are required to move the temperature control unit on the casters if the total weight of the
temperature control unit is
over 1.5 tons
.
▪
The parking brakes must be activated at the casters and/or the feet (if present) must be un-
screwed/activated before the temperature control unit is put into operation.
→
Page 30, section
»Unscrewing/activating the leveling feet (if any)«
2.1.3.2
Temperature control unit without casters
▪
An industrial truck must be used for positioning the temperature control unit.
▪
Do not move the temperature control unit alone.
▪
At least 2 persons
are required to move the temperature control unit.
▪
The industrial truck must have a lifting force equal to or greater than the weight of the tempera-
ture control unit. See the data sheet for the weight of the temperature control unit.
→
From page
▪
The leveling feet (if present) must be unscrewed/activated before the temperature control unit is
put into operation.
→
»Unscrewing/activating the leveling feet (if any)«
2.2
Unpacking
Starting up a damaged temperature control unit
DANGER TO LIFE FROM ELECTRIC SHOCK
Do not operate a damaged temperature control unit.
Please contact Customer Support.
→
PROCEDURE
Check for damage to the packaging. Damage can indicate material damage to the temperature
control unit.
Check for any transport damage when unpacking the temperature control unit.
Always contact your forwarding agent regarding the settlement of claims.
Observe the proper disposal of packaging material.
→
Page 15, section
2.3
Ambient conditions
Unsuitable ambient conditions / unsuitable installation
SERIOUS INJURY DUE TO CRUSHING
Comply with all requirements!
→
and
→
Page 26,
section
Make sure there is adequate fresh air available at the site for the circulation pump and the com-
pressors. The warm exhaust air must be able to escape upwards unhindered.
Free-standing models
For the connection data, see the data sheet.
→
Use of the temperature control unit is permitted only under normal ambient conditions in accord-
ance with the currently valid DIN EN 61010-1.