Function description
Chapter 3
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
39
Function
E-grade
Basic
E-grade
Exclusive
E-grade
Professi
-
onal
Remote control / data visualization via Spy software
X
X
X
E-grade evaluation versions available (valid for 30 days)
X
X
X
Storage/loading temperature control programs onto USB stick
–
X
X
Copying settings
–
–
X
Service data recorders (flight recorder)
X
X
X
PB commands
X
X
X
Communication Watchdog
–
–
X
Process data recording directly to a USB stick: Setpoint, Actual Value
Internal and Actual Value Process / Heating Capacity %, Cooling Capaci-
ty %, and Pump Pressure / Pump Speed, and VPC Pressure
–/–/–
X/X/–
X/X/X
Calendar start
–
–
X
3.5
Clock/event function
3.5.1
Rechargeable accumulator
“Pilot ONE” is fitted with a clock that continues to run even when the temperature control unit is
switched off. The energy required for this purpose is provided by a rechargeable accumulator, which
is automatically charged when the temperature control unit is switched on. The accumulator is
dimensioned so that the clock can also continue to run for prolonged switch-off intervals (up to
several months). If, after extremely prolonged switch-off time, time and date have been deleted,
leaving the temperature control unit switched on for a few hours will usually suffice (no tempera-
ture control required). During this time, you can reset the time and date.
If after switching the unit off and back on again, the previously set time and date re-appear, it can be
safely assumed that the rechargeable accumulator is defective. In this case, contact Customer Sup-
port. The telephone number can be found on page 84 in section
.
3.5.2
Programmable event function
The “Calendar Start” offers a programmable event function. This enables you to enter a time at
which the event is repeatedly triggered on a daily basis (until the activity in the menu is reset). 2
event types are currently selectable:
3.5.2.1
Event function “Alarm clock event”
Several acoustic signals are used.
3.5.2.2
Event function “Program event”
After selecting
“Program event”
when configuring the event function, you will be prompted for the
number of the program to be started. The program will be started automatically when the pro-
grammed event time is reached. If the temperature control unit is not active, this will also be start-
ed.
1
See Data Communication Manual. Anything controllable via the graphical user interface of the Pilot
ONE can be controlled by PB commands.