background image

4

General Purpose sling

Human Care is an ISO 13485:2016 certified 

Swedish medical device company. The Quality 

Management System is in compliance with US 

21 CFR part 820.
The product is CE marked in accordance with EU 

Directive 93/42/EEC (MDD) and/or EU Regula

-

tion 2017/745 (MDR), as class I medical device.
The CE mark is on the product.
This product has been tested in accordance 

with ISO 10535.

Compliances and Standards

Safety Instructions

• 

Check that the client is sitting securely in the 

sling before starting to lift him/her. 

• 

Check that the lift is balanced when moving, 

raising and/or lowering the client. If you 

are using a Human Care lift with telescopic 

hanger bar, it is important that the bar hooks 

are in the same position on both sides. 

• 

Our slings have been designed and tested 

for indoor use with one client at a time and 

must only be used for transferring a person. 

This product is not a swing.

•  Make sure no sharp objects are in contact 

with the sling. 

Incorrect

Correct

The process of lifting and moving a person 

always involves risks. Therefore, you must read 

the user manuals for the lift and the lifting 

equipment carefully. Always make sure that 

the lifting equipment you are using is intended 

for use with the lift. As a care provider, you are 

responsible for the client’s safety and you must 

know whether the client will be able to cope 

with the lifting process. The sling must be fitted 

to meet the client’s specific needs.  If you are in 

any doubt, please contact the supplier. 

• 

Carry out a careful visual inspection of the 

sling every time you use it.

•  Check that the right sling is being used for 

the client. The model, size and material must 

meet the client’s needs.

• 

Plan the moving and lifting process so that it 

is as safe as possible.

• 

Before raising the client, check that the sling 

is correctly attached to the lift/hanger bar 

once the lift straps are stretched.

•  It is important that the hooks of the hanger 

bar are in the correct position, with the 

opening of the hook facing upwards on both 

sides to securely hold the straps in place. 

Recheck the position of the hooks once the 

lift straps are stretched.

• 

It is important to always check that the loops 

are at the bottom of all hooks (see pictures) 

and securely placed below the locking pin on 

all hooks.

Summary of Contents for NSB-103

Page 1: ...User manual Bruksanvisning Manuel de l utilisateur Gebrauchsanweisung General purpose sling NSB 103 NSB 104 NSB 105 NSB 109...

Page 2: ...ompliances and Standards 4 Safety Instructions 4 General Description 5 Technical Specifications 5 Product Label 7 Suspension Alternatives 8 Hanger Bars for Human Care Lifts 9 Compatibility 9 Sling App...

Page 3: ...and ensure you fully understand them before using this product Important You must read the user manual for your device prior to usage Keep this booklet and information handy for future use Product mod...

Page 4: ...olves risks Therefore you must read the user manuals for the lift and the lifting equipment carefully Always make sure that the lifting equipment you are using is intended for use with the lift As a c...

Page 5: ...mm 17 318 kg 700 lbs General purpose sling reusable NSB 103 L Polyester net 8 straps L 1143 mm 45 1029 mm 41 533 mm 21 318 kg 700 lbs General purpose sling reusable NSB 103 XL Polyester net 8 straps X...

Page 6: ...lbs General purpose sling single patient use with clips NSB 104C XL polymethylpentene 4 straps XL 1334 mm 53 1143 mm 45 679 mm 27 318 kg 700 lbs General purpose sling reusable with clips NSB 103C S P...

Page 7: ...ot Bleach c Tumble Dry Low Heat d Do Not Dry Clean e Never Iron f Do Not Use Fabric Softener 18 Owner name For example write the client s name or another unique identifier 19 Indicate annual inspectio...

Page 8: ...section through the strap on the other leg section before attaching them to the hanger bar An individual assessment of the client s needs must always be carried out before choosing an alternative conn...

Page 9: ...er bars made by manufacturers other than Human Care the combined system needs to be designed with the same sling attachment method for the hanger bar of floor lifts and ceiling lifts meaning loop sty...

Page 10: ...aps please see page 2 There are diffe rent ways to attach the loops to a hanger bar and to position the legs This sling can be used both with 2 3 or 4 point hanger bars The image above shows attachmen...

Page 11: ...e the desired positioning of the legs 4 Lower the hanger bar and securely attach straps see page 2 with appropriate loop For sling with clips Attach the leg section lifting clips to the lifting frame...

Page 12: ...achieve the divided position bring the leg sections up between the thighs thread one leg strap through the bottom opening of other left and attach the leg straps to the hanger bar cross over style For...

Page 13: ...the seams and the lifting straps Damaged lifting equipment must not be used Cleaning Human Care slings are intended for individual use To avoid infections or transmission of bac teria it is important...

Page 14: ...Human Care slings are intended to be used with one patient at a time To avoid possible trans mission of infectious diseases it is important to clean the sling if it has become soiled or if it is to b...