APPLICATIONS

 The HUMASCOPE biological microscope is suited for use in medical and sanitary establishments, laboratories,

institutes, agricultural research networks, colleges and universities for clinical microscope inspections and may be

used as an instructional instrument. It can also be used in research work such as biology, bacteriology, cytology,

histology and pharmaceutical chemistry.

3

 

SPECIFICATIONS

 Ocular

Designation

Magnification

Field Of View Diameter

Focal Length

Wide Field

10X

18mm

24.94mm

Plan Field

16X

11mm

15.58mm

 Objective

Designation

Magnification

Numerical Aperture

Working Distance

4X

0.10

17.912mm

10X

0.25

2.04mm

40X

0.65

0.65mm

Achromatic

100X(Oil)

1.25

0.09mm

 Total Magnification

Objective

Total Magnification

Ocular

4X

10X

40X

100X

10X

40X

100X

400X

1000X

16X

64X

160X

640X

1600X

 

Abbe-type condenser N.A.=1.25 with an adjustable iris diaphragm

 

Coarse focussing range: 30mm with a coarse focussing stopper

 

Fine focussing range: 30mm

 

Fine focussing division interval: 0.002mm

 

Area of the stage: 124 X 153mm

 

Specimen shifting range: transversal 70mm, longitudinal 50mm

 

6V20W built-in variable-brightness lamp

 

5-objective revolver

 

Köhler illumination

Summary of Contents for HumaScope

Page 1: ...Gtl Rbnod User Manual Cat No 16100 1...

Page 2: ...i Qduhrhnm Khrs ne sgd L mt k No DATE Rev REVISION DESCRIPTION 1 01 2004 07 First edition 2 02 2006 08 Addition of Chapter 4 2 1 K hler Illumination Update Specifications...

Page 3: ...ii...

Page 4: ...llation of this product This warranty does not apply to damages incurred in shipment of goods Any damage so incurred shall be reported to the freight carrier for settlement or claim 3 INTENDED USE OF...

Page 5: ...at least potentially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the labo...

Page 6: ...Holder 4 4 4 Ocular Objective and Nosepiece 5 4 5 Coarse and Fine Focussing Knobs 5 4 6 Condenser 5 4 7 Immersion Oil 5 5 OPERATION 7 6 STANDARD ACCESSORIES 7 7 CARE AND MAINTENANCE 7 8 DISPOSAL MANA...

Page 7: ...2 9...

Page 8: ...trument It can also be used in research work such as biology bacteriology cytology histology and pharmaceutical chemistry 3 SPECIFICATIONS Ocular Designation Magnification Field Of View Diameter Focal...

Page 9: ...e nosepiece which holds the objective lenses to the 10X or 20X lens Open the field diaphragm as far as possible 2 Check if the specimen is illuminated It will help to place a piece of paper over the t...

Page 10: ...focus the stopper may be used to lock in the focus This also allows the stage to be lowered and raised again to exactly the same position When the position is no longer needed simply loosen the stopp...

Page 11: ...6 9 Figure 4 0 HUMASCOPE Light Source Mechanical Stage Revolving Nosepiece Ocular Objective Focussing Knobs Condenser...

Page 12: ...free location The instrument should be covered with dust cover after use 2 Never disassemble the lenses these are precisely adjusted and glued in place at the factory Stains can be removed from the le...

Page 13: ...ible Besides the optical advantages darkfield illumination is very beautiful and gives an almost science fiction like image What is the difference between darkfield and brightfield The difference betw...

Page 14: ...light source with proper wavelengths for excitation is also needed For normal fluorescence applications this is a mercury vapour arc burner HQVHV objective lens gathers light from the specimen ocular...

Page 15: ...HUMAN Gesellschaft f r Biochemica und Diagnostica mbH Max Planck Ring 21 65205 Wiesbaden Germany Tel 49 61 22 99 88 0 Fax 49 61 22 99 88 100 e Mail human human de www human de 02 2006 08...

Reviews: