Procedure:
1.
Make sure that the code number on
the code strip is the same as on the
glucose test strip vial.
2.
Insert a code strip into the test port
of the meter. A code number will appear
on the screen.
3.
Make sure the code number on the
screen is the same as the code number
on the code strip and on the test strip
vial. Remove the code strip. “ ” should
appear on the screen, indicating that the
meter has been successfully coded to the
respective test strip lot.
Caution:
If “E-E” appears on the screen during
the coding procedure, there may be a
problem with the code strip or a wrong
category code strip is being used. Check
that the correct strip type is being used.
Perform the coding procedure again,
using a different code strip if necessary.
If “E-E” continues to appear, please
contact your authorized distributor for
service.
2.4 Control Test
The purpose of HumaSens Glucose
control solution is to check the perfor-
mance of the HumaSens Glucose Meter.
The system is performing correctly if the
control test result falls within the indica-
ted control range listed on the test strip
vial. Only use HumaSens Glucose control
solution (
[REF]
17545).