11-2. Detailed Assembly Procedures
11-2-1. Installing the Bulb [TRF, TRF-P]
1)
Use only the designated bulb 12V100W Halogen lamp. (PHILIPS 7724)
2)
Fully loosen the lamp housing clamping screw ① of the lamp housing cover with provided Allen
screwdriver.
3)
Lift the lamp housing cover ② upward to remove.
4)
Holding the bulb with gloves or a piece of gauze, depress the bulb clamping lever ③ and insert bulb pin
fully into the pin holes ④ on the lamp socket. Gently release the bulb clamping lever to the original
position to secure the bulb.
5)
Side the lamp housing cover onto the housing base from the above. Tighten the clamping screw ①
while pressing downward on the cover.
Fig.21
➊
➋
Fig.22
➌
➍
The bulb, the lamp socket and areas near these will be extremely hot during and right
after use. Set the main switch off, disconnect the power cord from the wall outlet, then
wait until lamp housing has cooled down before replacing the bulb with a new of the
designated type.
To prevent reduced bulb life or cracking, do not touch the bulb with bare hands. If
fingerprints are accidently left on the bulb, wipe the bulb with a soft cloth.
When you dispose the device and accessory kit, Comply with related law and recycling plan.
Disposal of light bulb can cause the environmental pollution.
When you dispose the packing material, Comply with related law and recycling plan.