11-2-3. Attaching the Condenser [TRF, TRF-P]
1)
Turn the coarse adjustment knob ① to raise the stage to its upper limit.
2)
Turn the condenser height adjustment knob ② to lower the condenser holder to its lowest position.
3)
Fully loosen the condenser clamping knob ③.
4)
Position the condenser so that the aperture scale ④ is located in front.
5)
Tighten the condenser clamping knob ③ and gently raise the condenser holder to its height position.
11-2-4. Attaching the Stage
1)
Turn the coarse adjustment knob ① to lower the stage to its lower limit.
2)
Place the stage on the substage and tighten the stage clamping knob ②.
➊
➋
➌
➍
➊
➋
Fig.24
Fig.25