background image

16    SLIT LAMP HS- 7000------------------- ----------------------------------------------------------------

4.

Standard Components

The instrument is delivered in package.  Taking it out      the package,  make sure  that 

none of the following parts is missing.

A table top(N.B. the table top isn’t supplied  with the slit  lamp for twin table) on 

which  are  placed:    A  SMPS  box(15)  including  the  main  switch  &  a  warning 

light  (13),  Two  sliding  guides  for  the  base  with  orthogonal  movements(17),  A 

sliding plate for the position device(19), A drawer(16);

A base with orthogonal movements.

A  microscope  with  either  dual  magnification  or  a  system  (31)  including 

eyepieces.

An optical group of the slit projector.

A head rest module (59).

A microscope arm (45).

This manual.

A set of accessories including; two wheels for the  sliding  guides(18); a setting 

rod(43);  a  protective  cover(46);  a  spare  bulb  12V  30W       ogen  bulb(53);  two 

protection  fuses(48);  a  breath  screen(19);  a  chin-rest  paper(47);a  hole 

cover(45); a power cable; two patient’s handles(50)

l

l

l

l

l

l

l

l

Summary of Contents for HS-7000

Page 1: ...SLIT LAMP HS 7000 1 Operator s Manual SLIT LAMP MICROSCOPE HS 7000...

Page 2: ...rd For use of equipment in rated voltage less than 125Vac minimum 6A Type SJT or SVT 18 3AWG 10A max 3 0m long One end with Hospital Grade Type NEMA 5 15P other end with appliance coupler For use of e...

Page 3: ...ce and is not required to update this documentation to reflect such changes 2009 HUVITZ Co Ltd 689 3 Geumjeong dong Gunpo si Gyeonggi do 435 862 Republic of Korea All rights are reserved Under copyrig...

Page 4: ...sories 17 6 Composition 18 7 Installation 24 8 Trial Operations 28 9 Main Features 29 10 Ordinary Maintenance 30 1 1 Indication for use 1 2 Classification 6 2 1 Introduction 7 2 2 Safety Symbols 8 2 3...

Page 5: ...acturer s declaration electromagnetic emissions 34 11 2 Guidance and manufacturer s declaration electromagnetic immunity 35 11 3 Guidance and manufacturer s declaration electromagnetic immunity 37 11...

Page 6: ...lied parts Protection against harmful ingress of water Ordinary IPX0 Degree of safety in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide Not suitable for...

Page 7: ...nt and the user operator It is for this reason that the following safety notices have been placed appropriately within the text of this manual to highlight safety related information or information re...

Page 8: ...t has the certification symbol printed explicitly on the product label or sticker I and O on po wer switch represent ON and OFF respectively Type B Isolated patient connection This symbol identifies a...

Page 9: ...n Directive 2002 96 EC All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the l...

Page 10: ...t hand Where the equipment is exposed to direct sunlight Where the temperature changes extremely Normal operating temperature range is from 10 C to 40 C Humidity is from 30 to 75 Altitude is from 700...

Page 11: ...power cord into the outlet before the connection between devices of the equipment is completed This can generate the defect Pull out the power cord with holding the plug not the cord Avoid places wher...

Page 12: ...device in humid rooms Humidity should be maintained between 30 and 75 for normal operation Do not expose the device to water splashes dripping water or sprayed water Do not place containers containin...

Page 13: ...for persons operating and maintaining the equipment 11 Do not force cable connections If a cable does not connect easily be sure that the connector plug is appropriate for the receptacle socket If you...

Page 14: ...diagnostic procedures bio microscopic examination of the eye As long as the slit lamp is in its original package it can be exposed to the following environmental conditions for up to 15 weeks shipping...

Page 15: ...f the radiance and the exposure time If the value of radiance were reduc in half twice the time would be needed to reach the maximum exposure limit While no acute optical radiation hazards have been i...

Page 16: ...rthogonal movements 17 A sliding plate for the position device 19 A drawer 16 A base with orthogonal movements A microscope with either dual magnification or a system 31 including eyepieces An optical...

Page 17: ...lowing standard accessories are provided with the instrument protection cover 46 a setting rod 43 spare bulbs 12V 30W Halogen bulb 53 chin rest paper 47 spare fuses 250Vac T2 0AL a breath screen 5 pow...

Page 18: ...longitudinal and vertical movements x y z 10 Base with orthogonal movements 11 Shaped tabletop 12 Warning light indicating the ignition of the SMPS is on 13 Main switch Warning light 14 Luminosity sel...

Page 19: ...orizontal tilting 36 Control lever for vertical tilting 37 Microscope position stop 38 Graduated ring 90 0 90 determining the inclination of the slit during rotation 39 Lamp cover 40 fuse slots 41 Con...

Page 20: ...IT LAMP HS 7000 54 Bulb fastening knob 55 Mirror 56 Capture connector For the image system 57 Capture switch 58 Luminosity selector connector Connect the Luminosity selector to the SMPS 59 Head rest m...

Page 21: ...SLIT LAMP HS 7000 21 Figure 2 Composition I...

Page 22: ...22 SLIT LAMP HS 7000 Figure 3 Composition...

Page 23: ...SLIT LAMP HS 7000 23 Figure 4 Composition...

Page 24: ...ckaging to be able to stand up to all kinds of transport and standard storage conditions Please contact the suppliers if any irregularities are found in the appliance due to transport after unpacking...

Page 25: ...ides situated on the upper part of the table top checking that the wheels are in line Then Place the crankcase 59 and make sure that its holes fit together with those of the toothed guides 17 Tighten...

Page 26: ...crew driver Put the microscope 31 in the right position making sure that it is placed well in the stop 37 then tighten it with the knob 34 situate the right side of the microscope Fasten the breath sc...

Page 27: ...SLIT LAMP HS 7000 27 Figure 7 Assembling the instrument...

Page 28: ...and his forehead against the forehead rest Lift or lower the chin rest by moving the handle 25 so that the patient s eyes are in line with the notches on the chin rest 29 Switch the instruments on pr...

Page 29: ...s 6X 10X 16X 25X 40X 5 enlargements model Tilting 5 10 15 20 Microscopy of the back part and of the posterior vitreous body by means of the Hruby lens or a contact options Goldman lens Observation of...

Page 30: ...g lamps Please place lamps after cooling down The heated lamp may burn your fingers Don t touch any other parts of glass tube with the bare fingers Fingerprints if any must be wiped off from the glass...

Page 31: ...the pl tic envelope never touch the bulb whit fingers Fasten the bulb with the knob Make sure that the stems of the bulb are completely inserted insert the plug in the main supply socket Don t Touch t...

Page 32: ...e placed on the back of the SMPS and inserted power socket Before doing anything else disconnect the plu from the main supply socket Remove the burned out fuses Substitute the fuses and check Replace...

Page 33: ...ub the body and the probe once a month Do not use lacquer thinner ethylene or the oxidizing substance If the instrument is not in use dust must be kept away by covering the instrument with the supplie...

Page 34: ...F Emissions CISPR 11 Group 1 The EUT uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment R...

Page 35: ...kV for input output lines 2kV for power supply lines 1kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV differential...

Page 36: ...tible power supply or a battery Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels character ristic of a typical location in a typical...

Page 37: ...assure that it is used in such an environment Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80MHz 3 V m 80 MHz to 2 5GHz 3 Vrms 150 kHz to 80MHz 3 V m 80MHz to 2 5GHz Portable...

Page 38: ...people a Field strengths from fixed transmitters such as base tations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be p...

Page 39: ...frequency of transmitter m 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2 5GHz V1 3Vrms E1 3V m E1 3V m 0 01 0 116 0 1166 0 2333 0 1 0 368 0 3687 0 7378 1 1 166 1 1660 2 3333 10 3 687 3 6872 7 3785 100...

Page 40: ...5 Immunity and Compliance Level Immunity test IEC 60601 Test Level Actual Immunity Level Compliance Level Conducted RF IEC 61000 4 6 3Vrms 150kHz to 80MHz 3Vrms 3Vrms Radiated RF IEC 61000 4 3 3Vrms 8...

Page 41: ...Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80MHz 3 V m 80 MHz to 2 5GHz 3 Vrms 150 kHz to 80MHz 3 V m 80MHz to 2 5GHz The EUT must be used only in a shielded location with a minimum RF shielding eff...

Page 42: ...cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment...

Page 43: ...SLIT LAMP HS 7000 43 12 Electric Diagram...

Page 44: ...25x 40x 12 5x eyepiece or 5x 8x 12x 20x 32x 10x eyepiece Real field of view 38 5 24 0 15 0 9 0 6 0 mm Interpupillary adjust 55 80 mm Slit length 0 3 12 mm continuous Slit width 0 12 mm continuous Slit...

Page 45: ...Min 28mm Longitudinal Movement Min 78mm Lateral Movement Min 98mm Fine base Movement Min 10mm Power supply 100 240 V AC 50 60HZ 2 0A Power consumption 70VA Instrument Voltage 12V DC Halogen Projectio...

Page 46: ...e information such as Model and Serial Numbers To do so fill up the following table as soon as you purchase our product You can look up these information at any time The serial number is found on the...

Page 47: ...7 How to Contact HUVITZ Co Ltd Write us at Tel Fax URL http www huvitz com e mail hs support huvitz com HUVITZ Co Ltd Huvitz B D 689 3 Geumjeong dong Gunpo si Gyeonggi do South Korea 435 862 82 31 442...

Page 48: ...48 SLIT LAMP HS 7000 15 Particular Information 15 1 Relative spectral output graph 1 Halogen Lamp 2 Fixation Lamp...

Page 49: ...imum aperture CAUTION CAUTION The light emitted from this instrument is potentially hazardous The longer the duration of exposure the greater the risk of ocular damage Exposure to light from this inst...

Reviews: