background image

10

3

Operation: Extract 1 millilitre of pure sulfur hexafluoride with a

one-litre cylinder from the steel bottle. Inject it into 100-millilitre cylinder and
dilute it with air outdoors. The concentration degree is 10000ppm. Again,
extract 5 millilitre 10000ppm sulfur hexafluoride with a twenty-milliliter from a
100-millilitre cylinder and inject it into another 100-millilitre cylinder. Dilute it
with air outdoors to 100 millilitres. Its concentration degree is 500ppm. The
same operation can be applied for 100ppm, 10ppm, 1ppm, 0.1ppm and
0.01ppm. Each should be marked to be tested.

2. Demarcation of the leakage detector:
When the instrument is shut down, press OK to switch on the

detector( enter the interface of detection demarcation). After 15 minutes the

machine gets stable. Press

&

at the same time to enter

demarcation interface. At this time the demarcation number is shown on the
interface. Press + or

to alter the number of points which need to be

demarcated. Press OK and then demarcate the first point. Input OPPM and
then press OK. Input 1010, which is the first point

s collecting number. Press

OK and then QUIT to enter the demarcation interface. Thus the demarcation of
the first point is finished. The detection of the prepared SF6 gas begins. Use
20-millilitre cylinder to extract 10 millilitre of the prepared sulfur hexafluoride
with different concentration degree. Take off the needle of the cylinder and
connect the intake port of cylinder and probe. The probe will take in the gas to
be detected by itself. At the moment the signal value on the lower left corner of
the LCD of the detector will increase. Record the reading when the displayed
data turns stable ( Record the seven sample gases in turn) . The interface
shows the demarcation point now. press + or

to alter the number of points to

be demarcated and press OK. Then demarcate the first point (zero point).
Input OPPM and press OK. Input the collecting number 1010 at zero point and

then press OK. Press

again to alter the point number to the second

point( 0.01 PPM ) and then press OK again to input the collecting value of 0.01

Summary of Contents for HTJL-H

Page 1: ...al You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may...

Page 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Page 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Page 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Page 5: ...5 Contents Warning 6 I Product introduction 6 II Technical data 7 III Product structure 7 IV Operation procedure and method 8 V Adjustment method 9 VI Announcements 11...

Page 6: ...The concentration degree of sulfur hexafluoride can be displayed by a detection of the degree of ionization This instrument is a small vacuum pump with its searching unit and main engine separated Tho...

Page 7: ...ppm Volume ratio Pumping speed of vacuum pump 0 5 litre second Length of the vacuum tube 4 m Continual working time not more than 3 hours Working condition alternating current power supply 220V 50Hz E...

Page 8: ...e concentration of SF6 gas detected in real time so it is unnecessary to check the table again The small value below is the base value which is only used as a reference for the stability of the instru...

Page 9: ...m value using The key changes When the volume ratio is greater than the alarm point the device will emit a beeping alarm V Adjustment method 1 Making up different concentration degree sulfur hexafluor...

Page 10: ...to be demarcated Press OK and then demarcate the first point Input OPPM and then press OK Input 1010 which is the first point s collecting number Press OK and then QUIT to enter the demarcation interf...

Page 11: ...ppropriate amount The three adjustable potentiometer inside the searching unit has been set for its output voltage and signal output value before selling In generall it doesn t need to be adjusted aga...

Reviews: