11

PPM sample gas. Press OK and then

to go to the next point. In this way

the demarcation of all the seven points is finished ( six sample gas value and
one zero point value ). After all the sample gas

s demarcation is finished, the

user can restart the leakage detector for the next detection.

VI. Announcements

1. Please keep the body of the whole instrument clean and tidy. Avoid the

entrance of any impurities to enter the probe or the vacuum pump. The
machine should be kept in a dry and windy place to avoid humidity.

2. Since the probe needle valve has been adjusted well before it is sold,

do not adjust it in normal condition. If the place of the needle valve changed
due to any improper use or uncertain reason, the user should readjust it till the
intaking gas reaches an appropriate amount. The three adjustable
potentiometer inside the searching unit has been set for its output voltage and
signal output value before selling. In generall it doesn

t need to be adjusted

again. If there is a need to adjust it, please refer to the technical parameter of
the instrument to do it, or the inside component may be burnt. Each time when
the inside potentiometer is adjusted, the instrument has to be demarcated too.
Or else the detectiong data will not be accurate.

3. After a long-term use the O-like rubber washer at the connection point

of the needle valve gets aged. As a result, the leakage at the connection point
affects the vacuum degree. So it should be changed at once.

44. Do not overturn the detector or shake it violently. It may cause

abnormal state of the searching unit or other movable components and affects
the detector

s function.

5. If there is no signal or a snowflake screen when the detector is started,

press

and then QUIT to refresh it.

Summary of Contents for HTJL-H

Page 1: ...al You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may...

Page 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Page 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Page 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Page 5: ...5 Contents Warning 6 I Product introduction 6 II Technical data 7 III Product structure 7 IV Operation procedure and method 8 V Adjustment method 9 VI Announcements 11...

Page 6: ...The concentration degree of sulfur hexafluoride can be displayed by a detection of the degree of ionization This instrument is a small vacuum pump with its searching unit and main engine separated Tho...

Page 7: ...ppm Volume ratio Pumping speed of vacuum pump 0 5 litre second Length of the vacuum tube 4 m Continual working time not more than 3 hours Working condition alternating current power supply 220V 50Hz E...

Page 8: ...e concentration of SF6 gas detected in real time so it is unnecessary to check the table again The small value below is the base value which is only used as a reference for the stability of the instru...

Page 9: ...m value using The key changes When the volume ratio is greater than the alarm point the device will emit a beeping alarm V Adjustment method 1 Making up different concentration degree sulfur hexafluor...

Page 10: ...to be demarcated Press OK and then demarcate the first point Input OPPM and then press OK Input 1010 which is the first point s collecting number Press OK and then QUIT to enter the demarcation interf...

Page 11: ...ppropriate amount The three adjustable potentiometer inside the searching unit has been set for its output voltage and signal output value before selling In generall it doesn t need to be adjusted aga...

Reviews: