background image

6

Warning:

The user has to confirm whether the vacuum pump has been filled with oil

before any use of the instrument except there is a special announcement.
Please add the oil to the pump or it will be damaged. Tear down the outer
cover of the main engine and refer to the instruction of the vaccum pump to
add the oil.

When the oil is added, the instrument cannot be overturned in the process

of transporting.

I. Product introduction

HTJL-H SF6 Gas leak Detector’s working principle is that when the

concentration of sulfur hexafluoride in the air changes its degree of ionization
under an electromagnetic field of high frequency varies at the same time. The
concentration degree of sulfur hexafluoride can be displayed by a detection of
the degree of ionization.

This instrument is a small vacuum pump with its searching unit and main

engine separated. Though small and light, it has a high sensitivity and stability.
Moreover, it responds fast and produce no toxic gas in its component(esp. the
searching unit).

This instrument has a new function of alarming setting. When the gas to

be detected has a concentration degree more than or equal to the
pre-determined value, it alarms automatically.

This instrument is used exclusively in the detection of sulfur hexafluoride

equipment or volume’s leakage by the industry of electricity, railway, electronic
production, chemical engineering, fire equipment and atomic physics research
institute, etc. It can do qualitative and quantitative detection quickly and
accurately.

Summary of Contents for HTJL-H

Page 1: ...al You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may...

Page 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Page 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Page 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Page 5: ...5 Contents Warning 6 I Product introduction 6 II Technical data 7 III Product structure 7 IV Operation procedure and method 8 V Adjustment method 9 VI Announcements 11...

Page 6: ...The concentration degree of sulfur hexafluoride can be displayed by a detection of the degree of ionization This instrument is a small vacuum pump with its searching unit and main engine separated Tho...

Page 7: ...ppm Volume ratio Pumping speed of vacuum pump 0 5 litre second Length of the vacuum tube 4 m Continual working time not more than 3 hours Working condition alternating current power supply 220V 50Hz E...

Page 8: ...e concentration of SF6 gas detected in real time so it is unnecessary to check the table again The small value below is the base value which is only used as a reference for the stability of the instru...

Page 9: ...m value using The key changes When the volume ratio is greater than the alarm point the device will emit a beeping alarm V Adjustment method 1 Making up different concentration degree sulfur hexafluor...

Page 10: ...to be demarcated Press OK and then demarcate the first point Input OPPM and then press OK Input 1010 which is the first point s collecting number Press OK and then QUIT to enter the demarcation interf...

Page 11: ...ppropriate amount The three adjustable potentiometer inside the searching unit has been set for its output voltage and signal output value before selling In generall it doesn t need to be adjusted aga...