Guidance and manufacturer’s declaration – electromagnetic immunity 

The Hybresis Charging Station is intended for use in the electromagnetic environment specified below. The 

customer or the user of the Hybresis Charging Station should assure that it is used in such an environment. 

Immunity test 

IEC 60601 

test level 

Compliance level 

Electromagnetic 

environment - guidance

Electrostatic discharge 

(ESD) 
 

IEC 61000-4-2 

±6kV contact 

 
±8kV air 

±6kV contact 

 
±8kV air 

Floors should be wood, 

concrete or ceramic tile. If 
floors are covered with 

synthetic material, the 
relative humidity should 

be at least 30%. 

Electrical fast 
transient/burst 

 
IEC 61000-4-4 

±2kV for power supply 
lines 

 
±1kV for input/output 

lines 

±2kV for power supply 
lines 

 
±1kV for input/output 

lines 

Mains power quality 
should be that of a typical 

commercial or hospital 
environment. 

Surge 
 

IEC 61000-4-5 

±1kV differential mode 
 

±2kV common mode 

±1kV differential mode 
 

±2kV common mode 

Mains power quality 
should be that of a typical 

commercial or hospital 
environment. 

Voltage dips, short 
interruptions and voltage 

variations on power 
supply input lines 

 
IEC 61000-4-11 

<5% U

T

(>95% dip in U

T

) for 0,5 

cycle 
 

40% U

T

(60% dip in U

T

) for 5 

cycles 
 

70% U

T

(30% dip in U

T

) for 25 

cycles 

 

<5% U

T

(>95% dip in U

T

) for 5 sec 

<5% U

T

(>95% dip in U

T

) for 0,5 

cycle 
 

40% U

T

(60% dip in U

T

) for 5 

cycles 
 

70% U

T

(30% dip in U

T

) for 25 

cycles 

 

<5% U

T

(>95% dip in U

T

) for 5 sec 

Mains power quality 
should be that of a typical 

commercial or hospital 
environment. If the user 

of the Hybresis Charging 
Station requires 

continued operation 
during power mains 

interruptions, it is 
recommended that the 

Hybresis Charging Station 
be powered from an 

uninterrupted power 
supply or a battery. 

 

Power frequency 
(50/60Hz) magnetic field 

 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency 
magnetic fields should be 

at levels characteristic of 
a typical location in a 

typical commercial or 
hospital environment. 

NOTE U

T

 is the a.c mains voltage prior to application of the test level. 

 

6

Summary of Contents for 199586

Page 1: ...READ THE HYBRESIS CONTROLLER AND PATCH INSTRUCTIONS FOR USE FOR ADDITIONAL IMPORTANT INFORMATION REF 199586 Charging Station Instructions For Use ...

Page 2: ...ment not to dispose of WEEE as municipal waste Contact your local distributor for information regarding disposal of the unit and accessories Dangerous Voltage Underwriters Laboratories Inc indicates product meets US and Canadian product safety standards This device complies with UL 60601 1 and CSA C22 2 No 601 1 M90 Precautionary Instructions Keep Dry Keep Away from Sunlight Class II Equipment For...

Page 3: ...an unshielded manner Portable and mobile RF communications equipment can affect Medical Electrical Equipment This unit generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation Harmf...

Page 4: ...harging Station charges up to four Hybresis Controllers maintains them in a fully charged state and provides a convenient place for storage General When a Controller is inserted into one of the charging bays the Charging Station will automatically detect the battery level of the Controller and will begin to charge it 2 ...

Page 5: ...ntroller battery not charging properly Return Controller to Empi for service Charging Station Lights The two windows on each of the charging bays indicate the following On rare occasions the problem may be due to a defective Charger If you have another Controller please charge it to determine if the Charger is functioning correctly or not If the same problem occurs with multiple Controllers then t...

Page 6: ... maximum Environmental Storage Temperature 5 40 C 41 104 F Conditions DO NOT expose to temperatures above 50 C 122 F Operating Temperature 15 30 C 59 86 F Ingress of Not protected against ingress Water of water IPX0 Dimensions 8 5 cm H x 15 cm W x 12 5 cm L Weight 421g Class II Continuous Operation 4 ...

Page 7: ...ISPR 11 Group 1 The Hybresis Charging Station uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class A The Hybresis Charging Station is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage p...

Page 8: ... 2kV common mode 1kV differential mode 2kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT ...

Page 9: ...ransmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Page 10: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 3 5 P V1 80 MHz to 800 MHz d 3 5 P E1 800 MHz to 2 5 GHz d 7 P E1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation a...

Page 11: ...When returning any products please include your name address phone number and a description of the problem Return to Empi Attn Repair Department 47492 SD Hwy 22 Clear Lake SD 57226 Repair and Maintenance There are no user serviceable parts inside the Charging Station If the station appears to be non functional contact your Authorized Empi Distributor or contact Empi directly Clean the case as need...

Page 12: ...on 199587 Hybresis Controller 199589 Hybresis Patch Limited Warranty I Warning While in the opinion of Empi Empi the use of the Hybresis System the Product has met with some success Empi makes no warranties to the purchaser as to the effectiveness of the product 10 ...

Page 13: ...e or during use III Limitation of Liabilities and Disclaimer ofWarranties A Empi s sole obligation in the case of any breach of its warranties set forth in Paragraph IIA above shall be at Empi s option to repair or replace the Product without charge to Purchaser or to refund the purchase price of the Product In order to recover under this Warranty Purchaser must send Empi written notice of the def...

Page 14: ...nsurance prepaid as soon as reasonably possible following receipt of the Product by Empi If Empi determines in its sole reasonable discretion that the Product does not contain defective workmanship or materials Empi will return the Product to the Purchaser freight and insurance billed to the Purchaser B This Warranty is voided immediately as to any Product which has been repaired or modified by an...

Page 15: ...s so the above limitation may not apply to the Purchaser D EMPI SHALL NOT BE LIABLE TO ANY PERSON FOR ANY DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES LOST PROFITS OR MEDICAL EXPENSES CAUSED BY ANY DEFECT FAILURE MALFUNCTION OR OTHERWISE OF THE PRODUCT REGARDLESS OF THE FORM IN WHICH ANY LEGAL OR EQUITABLE ACTION MAY BE BROUGHT AGAINST EMPI E G CONTRACT NEGLIGENCE OR OTHERWISE THE R...

Page 16: ...205 Hwy 22 East Clear Lake SD 57226 651 415 9000 800 328 2536 360354 Rev C 2007 2010 Empi 1 10 ...

Reviews: