Filtromat – OF5 C
4.6
Measuring the cleanliness code
When the OF5 C is switched on using the main switch of the control cabinet, the
FCU 2110-4 is switched on at the same time.
After a short time, the display of the FCU 2110-4 shows:
OK/Start measurement
Cancel Stop
Press the OK/Start key of the FCU. The measured flow rate appears on the display.
Now make the fine adjustment of the flow rate control to (typical case) approx. 100
+/50 ml/min.
If air is also drawn in during the suction process of conditioning unit KE 1310, it must
be vented to avoid distorting the measurement results.
It can take several minutes until the air has been removed from the system and the
flow rate has stabilized.
For a more detailed explanation of the special control menus available for controlling
the filter unit through the FluidControl Unit, see the operating instructions of the
FCU.