Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 95 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
16. No defined value from CS
Cause:
The charge pump for the CS runs for 2 minutes in
Automatic mode.
The CS has been reporting "no measured value from
CS" for 4 minutes.
The message is shown on the CS display. The FAM
continues to run. The CS is switched off for 15 seconds,
after which a new measurement is started.
Remedy:
If the measurement has failed 3 times in the "Filter to"
operating mode, then the message "Start screen, RH+P,
T and CS" will appear on the display.