Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 114 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
37. Saturation oil >= 90%. Measurement with CS 1xxx not possible!
Cause: Error-free
measurement
of contamination is not possible
with a water content
≥
90%. The charge pump of the CS is
switched off.
If you selected an operating mode for which the CS
controls the FAM, then the FAM will automatically jump
into a similar operating mode without control by the CS.
Example: D Filter to => D Filter
Remedy:
The display flashes and shows "Operating mode
changed".
It will not be possible to reset this error message until the
relative humidity is <90%.
It will not be possible to select operating modes in which
the CS controls the FAM until the relative humidity is
<90%.