Measured values
HMG 4000
Edition 2016-03-08 - V01 R04
[24]
Part No.: 669950
Using the "
Progression
" opens a
new window.
Here, the current history of the
measured values is shown as an
ongoing graphic.
(Also see chapter 7.2.5 "Display
progression")
7.1.5 Resetting measured value
zero point
You can re-specify the zero point via
the
button.
This function can be used, for
example, in order to hide any
pressure heads in a system. In case
a pressure transducer still has, for
example, 3 bar of pressure even
when the system is switched off,
this value can be used as the new
zero point for the display.
Using the
button, you can
delete the zero shift.
Re-zeroing can be done at any point throughout the measurement range of a
measurement channel.