Measured values
HMG 4000
Edition 2016-03-08 - V01 R04
[35]
Part No.: 669950
7.4.2 Digital state
The measured value corresponds to
the digital state at the input. It is
updated every 0.1 ms.
The measurement range goes from
0 to 1 and does not have any unit or
decimal points.
7.4.3 Frequency, drive speed,
flow rate
The frequency is measured in "
Hz"
and, depending on the setting, can
be converted to speed or flow rate
(1 Hz to 30 kHz).
With "
l/min
" and "
USG/min
" the "
K
Factor"
must also be entered. This
indicates the connection between
pulses and volume ("
Pulse/l"
or
"
Pulse/USG"
)
Frequencies under 1 Hz can be
measured with "
Advanced
measurement
".
7.4.4 Pulse count
The pulses are counted here. The
measured value corresponds to the
counter reading.
The measured value is updated
every ms. Up to 2 billion pulses can
be counted.