Hydro-Logic
®
Flexi Logger 300Ex
User Manual
22 August 2018
Page 10 of 74
Sensors permissible values
Any contact closure or pulse sensor (digital) or two wire analogue sensors may
be connected to the Flexi Logger 300Ex provided the sensor meets the following
criteria:
•
The sensor must have ATEX or IECEx certification for use in the
hazardous zone they are being used in.
•
The sensors output parameters are not greater than the maximum
permissible values detailed below:
Parameter
Sensor connector
permissible values
Voltage Out - Uo
22.7v
Current Out - Io
297mA at 14.6v*
Power Out - Po
1.1W *
Capacitance - Ci
0.2uF
Inductance - Li
1.61mH
*Combined total for all three connectors