Hydro-Logic
®
Flexi Logger 300Ex
User Manual
22 August 2018
Page 27 of 74
•
Tighten the gland gently using a wrench
– finger-tightened glands are not
tight enough.
When making up the connectors, ensure that the connector bodies are correctly
threaded, and the cable seals are tight. Incorrectly assembled connectors will
allow water in, which can permanently damage the sensor.
Vented cables
Cables from pressure-based water-level sensors are usually ventilated to allow
for barometric pressure correction. A thin vent tube runs the length of the cable
interior.
The Hydro-Logic
®
Flexi Logger 300Ex sensor connectors are vented to
atmosphere via a GORE
®
protective membrane. The membrane exposes the
sensor vent tube to atmospheric moisture, while blocking liquid water. The sensor
connector has loose terminal pins which allow barometric pressure to pass from
the sensor vent tube, through both mating connectors and out through the
GORE
®
membrane. There is no need to drill the sensor connectors.
When installing, trim the vent tube and unused conductors to about 10mm
beyond the end of the outer insulation. Do not crush the end of the vent tube.
Approved sensors
Hydro International have a list of approved sensors which can be supplied with
the Hydro-Logic
®
information.
If you have a sensor in one of the above categories which is not on the Hydro
International list, it may still be possible to use it. Please contact
to discuss support for a specific sensor.
Pre-wiring and configuration service
Experienced technicians will pre-wire and configure your sensors for you and
ensure measurements are transferred successfully to Timeview if you have
ordered this package.
This service
does not
include advanced alarm configuration, other setup or
system design services. Contact
Summary of Contents for Hydro-Logic Flexi Logger 300Ex
Page 1: ...www hydro int com LoggerSupport Hydro Logic Flexi Logger 300Ex User Manual ...
Page 66: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 66 of 74 ...
Page 67: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 67 of 74 ...
Page 68: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 68 of 74 ...
Page 69: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 69 of 74 ...
Page 71: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 71 of 74 ...
Page 72: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 72 of 74 ...