This may be reversed by first entering the **** password and then entering a new password.
For example :
PASSWD=****<CR> Enable
access.
PASSWD=BOMM<CR>
Save BOMM as the new password.
If the MiniLog is connected to a modem, then it may have the comms cable permanently
connected. In this situation, once the password is entered, access is permanently granted. The
“Bye” (
BYE
) command forces the MiniLog to cancel the password access, and also performs
the CLEAR1/2/3/4 sequence of commands. (When the sleep timer expires, password access is
also cancelled.)
The “Sleep” (
SLEEP
) command provides a timer, so that a “Bye” is performed if there is no
communications in the preset time. The timer is set in minutes. For example, if “sleep” is set
to 5 minutes, and there is no communications for a 5 minute period, then password access
will be cancelled, and the CLEAR1/2/3/4 will be performed. When Sleep is set to 0, the
function is disabled.
SLEEP = 5<CR>
Sets the sleep timer to 5 minutes (Acceptable values are 0 to 9)
The “Reset” (
RESET
) command performs a hardware reset, but does not affect the time,
date, parameters or logged data.
RESET<CR>
Perform a reset of the MiniLog. Note that this command is
equivalent to switching the logger off and then on again.
The “Version”(
VER
) command allows the MiniLog firmware version to be displayed.
VER<CR>
Displays
the
MiniLog firmware version and date.