Hydrological Services Pty Ltd
Data Logger Model ML1-420
©
Copyright
ML1-420 100-23
Issue 1.00 6 Apr, 2011
4.1.10 Status Window (ST)
The Status Window (ST) returns a summary of station set-up details together with a report of
the present conditions. Each set-up parameter is identified by name followed by its unique
command keyword, (parameter abbreviation). The Status window provides a 'plain English'
response for visual interpretation when the ST command is issued.
Command Example
ST
Site(SI):
Sydney
Rainfall Channel ID (CHID):
0012345
LoggerID(ID): MJ110001
Rainfall
Log
Start: 22/01/11
Date(D):
28/03/11
Rainfall Inc. (INC):
0.2/01
Time(T):
10:06:46
Rainfall Today (TDAY):
5.2
Baud Rate (BAUD):
9600
Rainfall Total (TOT):
1260.7
Response(RE): On
Rainfall
Units
(UD):
mm
Event Output (EV):
On
Events Left:
421763
River Channel ID (CHIDRV):
0000002
Memory Wrap (MW):
Off
River Log Start:
22/01/11
Go to sleep after (SLEEP):
5 min
River Last (WL):
3.726
Logging On/Off (LOG):
On
River Units (U420):
m
S/W Revision (VER):
1.00
River Height Ref (RvHR)(offset):
0.417
River Trans Ref (TxR)(range):
10.000
Clear1:
+++~~
River Trans.Int (TXI)mmm/tt:
015/05
Clear2: ATH^M/OK/9
Clear3:
Temperature
(TEMPC):
24.3C
Clear4:
BV+T Log Interval (BVInt)mmm:
060
Battery Volt (BV):
3.54/12.6V
--------------------------------------------------------------------------------------------------------------------------
NOTE : The words “
Rainfall”
and
“River
” in the above Status Window, is the result of the
TYPE parameter. This string can be changed to customise the Status Window to your
requirements.
Note that the data in the Status Window is grouped to describe each area of the ML1-20, that
is “general configuration” on the left, “digital input configuration” at the top right, “4-20mA
configuration” at the centre right, and “battery / temperature configuration” at the lower right.