Hydro-View / Hydro-Hub User Guide HD0864 Rev 1.0.0 65
Figure 104: Collection Bucket
6. The Average Unscaled for all available measurement modes is displayed on the
calibration page.
Figure 105: Average Unscaled Values Taken During Averaging
7. Add the Unscaled values to the chart by clicking ‘Add Row’.
Figure 106: Add Calibration Row
Figure 107: Average Unscaled Values Added to Chart
8. Thoroughly mix the collected material to ensure the moisture is evenly distributed.
9. Weigh a clean empty heat proof bowl.
150.00 g
Figure 108: Clean Bowl
10. Place a minimum of 500g of the material in the bowl. All other material must remain in
the airtight container until required.