***** Channel 1 *****
Variable : p1 (bar)
Sens.type: 0-20 mA
Measure Range
from : 0.000
to : 300.0
Zero point 0.000
[
▼
] [
▲
] Highlight „Variable“ [Enter].
[
▼
] [
▲
] Select the desired variable and units [Enter].
[
▼
] [
▲
] Highlight „Sens.type“ [Enter].
[
▼
] [
▲
] Select the output signal of the sensor [Enter].
[
▼
] [
▲
] Highlight „from“ [Enter].
Enter the lower limit of the measuring range of the sensor, e.g. [ 0 ] [Enter].
[
▼
] [
▲
] Highlight „to“ [Enter].
Enter the upper limit of the measuring range of the sensor,
e.g. [ 3 ] [ 0 ] [ 0 ] [Enter].
[
▼
] [
▲
] Highlight „Zero point“ [Enter].
If you want to execute a zero point equalisation, you have to be sure that the
sensor receives the value „0“, e.g. that there is no pressure. Then press [Enter].
Horizontal dashes are displayed in this line of the display, while the equalisation
is in execution. Then the equalised value will be displayed.
Press [ C ].
7.3.2 Confi gure Digital Input Channel (Frequency / K3)
Press [Menu].
[
▼
] [
▲
] Highlight „Channels“ [Enter].
[
▼
] [
▲
] Highlight „C3“ [Enter]:
***** Channel 3 *****
Variable : Q1 (l/min)
Value : 100.
[
▼
] [
▲
] Highlight „Variable“ [Enter].
[
▼
] [
▲
] Select the desired variable and units [Enter].
[
▼
] [
▲
] Highlight „Value“ [Enter].
[
▼
] [
▲
] Enter the calibration value of the sensor, e.g. [ 1 ] [ 0 ] [ 0 ] [Enter].
Press [ C ].
The calibration value is contained in the documentation of the sensor.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
1.
2.
3.
4.
5.
6.
7.
8.
MultiHandy 3020
Operating Instructions
14