Adding SDI-12 Devices via Address Scan
An SDI-12 device can be detected and setup automatically by using the Address Scan feature of iLink. The device(s) needs
to be attached to the iRIS and have each have a unique address.
A device is initiated by click on the [
START ADDRESS SCAN
] button in the
SDI-12 DEVICE
branch of the configuration tree.
The scan will start immediately and scans the address range 0-9, a-z, A-Z. To speed up the process, click the [
STOP SCAN
]
button when the
TOTAL DEVICES FOUND
is equal to the amount of device attached (already configured or not).
Once the scan is complete or it has been stopped by the user. The devices found on the bus will be displayed in the list at
the bottom of the scan window.
Select the devices to be added to the iRIS configuration by checking the
SELECT
box and clicking the [
FINISH
] button. The
devices selected will be added to the SDI-12 Devices branch of the configuration.