S902/3
4.
Release the calibration button when the
totaliser display shows
VER
followed by
the version number on the lower display.
5. The totaliser display will show the
following in a looping sequence:
CH 1
- Single channel pulser connected
CH 2
- Two channel pulser connected
REED -
Reed switch pulser connected
Press calibration button once when
desired option is displayed.
Select
REED
for AdBlue
pedestal.
6. The totaliser display will show the
following in a looping sequence:
1 DP
- One decimal place on display
2 DP
- Two decimal places on display
Press calibration button once when
desired option is displayed.
7. The totaliser display will show the
following in a looping sequence:
RELAY 10
- Ten pulse per litre reed
relay output.
OPTO 10
- Ten pulse per litre opto-
isolated output.
OPTO 100
- One hundred pulse per litre
opto-isolated output.
Press calibration button once when
desired option is displayed.