32
-
Step 4.
Let stand for one minute. To kill TB,
let stand for 2 minutes at room
temperature (20-25
℃
). Wipe dry
or allow to air dry. Dispose of
towel.
A cleaning and disinfection cycle refers to one cleaning
step and one disinfecting step. The meter has been
validated for 260 cleaning and disinfection cycles, which
is equivalent to cleaning and disinfecting once per week
over the 5 year use life of the meter. The meter has been
further validated for a cleaning step after each use.
If you notice any of signs of deterioration of physical
characteristics (e.g., polymer crazing (thin silver streaks
appear), cracking, swelling, dissolving, softening,
becoming brittle) or performance (e.g., power
malfunction, icon illegible, display or number unclear),
stop using the meter and contact Alliance International
customer service.
For technical assistance or questions about cleaning and
disinfection including any possible deterioration in the
meter appearance or performance, please call customer
service for assistance.