Specifications & Operating Ranges
For blood glucose testing
For blood
β
-ketone testing
Calibration
Plasma-equivalent
Assay method
Electrochemical
Battery life
1,000 tests
Power
Two 3.0 V lithium batteries
(disposable, type CR2032)
Memory
1,000 test results
Size
106 x 58 x 17 (mm)
Weight
72.6 g (with batteries)
Bluetooth® Smart
• Frequency range: 2.4–2.4835 GHz
technology
• Operating range distance: maximum
10 metres (unobstructed)
• Operating channels: 40 channels
• Security encryption: 128-bit AES
(Advanced encryption standard)
Relative humidity
10–90%
Measurement range
1.1–33.3 mmol/L
Sample size
Minimum 0.4 μL
Test time
5 seconds
Sample type
• Fresh capillary whole blood
• Fresh venous whole blood
(healthcare professionals only)
Temperature
5–45˚C (41–113˚F)
Hematocrit
15–65%
Measurement range
0.1–8.0 mmol/L
Sample size
Minimum 0.5 μL
Test time
8 seconds
Sample type
Fresh capillary whole blood
Temperature
15–30˚C (59–86˚F)
Hematocrit
30–60%