Test Range: 20-600 mg/dL (1.1-33.3 mmol/L)
•
Sample Size: Minimum 0.5 μL
•
Test Time: 5 seconds
•
Test Sample: Fresh capillary whole blood
•
Calibration: Plasma-equivalent
•
Assay Method: Electrochemical
•
Meter Test: Check Strip
•
Power: One (1) 3.0 V lithium battery
(non-rechargeable, replaceable, type CR2032)
•
Battery Life: 1,000 tests
•
Memory: 250 blood glucose values
•
Dimension: 95 X 42 X 18 (mm)
•
Weight: 42.4 g (with battery)
•
Operating Range
- Temperature: 10-40˚C/50-104˚F
- Relative Humidity: 10-90%
- Hematocrit: 20-60%
Specifications
Message
What It Means
What To Do
A used test strip
was inserted.
Repeat the test with
a new test strip.
The blood or control
solution sample was
applied before the
symbol
appeared.
Repeat the test with
a new test strip and
wait until the
symbol appears
before applying the
blood or control
solution sample.
The temperature
during the test was
above or below the
operating range.
Move to an area
where the
temperature is within
the operating range
(10-40°C/50-104°F)
and repeat the test
after the meter and
test strips have
reached a
temperature within
the operating range.